| Literature DB >> 24121997 |
Wooseok Jung1, Jungyoup Han, Junhai Kai, Ji-Youn Lim, Donggeun Sul, Chong H Ahn.
Abstract
A new sample-to-answer polymer lab-on-a-chip, which can perform immunoassay with minimum user intervention through on-chip reservoirs for reagents and single-channel assay system, has been designed, developed and successfully characterized as a point-of-care testing (POCT) cartridge for the detection of thyroid stimulating hormone (TSH). Test results were obtained within 30 minutes after a sample was dropped into the POCT cartridge. The analyzed results of TSH showed a linear range of up to 55 μIU mL(-1) with the limit of detection (LOD) of 1.9 μIU mL(-1) at the signal-to-noise ratio (SNR) of 3. The reagents stored in the on-chip reservoirs maintained more than 97% of their initial volume for 120 days of storage time while the detection antibody retained its activity above 98% for 120 days. The sample-to-answer polymer lab-on-a-chip developed in this work using the mass-producible and low-cost polymer is well suited for the point-of-care testing of rapid in vitro diagnostics (IVD) of TSH.Entities:
Mesh:
Substances:
Year: 2013 PMID: 24121997 DOI: 10.1039/c3lc50403d
Source DB: PubMed Journal: Lab Chip ISSN: 1473-0189 Impact factor: 6.799