A cross-calibration study of the GE-Lunar iDXA and prodigy for the assessment of lumbar spine and total hip bone parameters via three statistical methods.

Karen Hind1, William Cooper2, Brian Oldroyd2, Andrew Davies2, Laura Rhodes2.   

Abstract

This study assessed agreement between the GE Lunar iDXA and Prodigy densitometers for bone measurements and used 3 statistical methods to derive cross-calibration equations: linear regression, the Deming method, and multivariate analysis. Compatibility of machines for the measurements of bone mineral density, bone mineral content, and bone area also was explored. Eighty-five adults, age: 45.5 (SD 12.8) years; body mass index: 25.6 (SD 3.7) kg.m(-2) were measured once at the lumbar spine: L1-L4 and total hip on each densitometer, within 24 hours. Both linear regression and Deming analysis indicated that cross-calibration equations were required at the lumbar spine and total hip but not at the femoral neck. Multivariate analysis identified femur thickness and femur percent fat as predictive variables at the femoral neck and total hip. Bland Altman analysis (Prodigy-iDXA) indicated significant positive bone mineral density bias at the lumbar spine and femoral neck. Significant bone mineral content biases were observed at all 3 sites and bone area biases at both hip sites. These initial results suggest there are small significant differences in the bone parameters and that all 3 bone parameters should be evaluated when comparing densitometers, especially when there are differences in pixel size between the densitometers.
Copyright © 2015 The International Society for Clinical Densitometry. Published by Elsevier Inc. All rights reserved.

Keywords:  Agreement; DXA; analysis; bone; cross calibration

Mesh:

Year:  2013        PMID: 24120577     DOI: 10.1016/j.jocd.2013.09.011

Source DB:  PubMed          Journal:  J Clin Densitom        ISSN: 1094-6950            Impact factor:   2.617


  4 in total

1.  Cross-Calibration of GE Healthcare Lunar Prodigy and iDXA Dual-Energy X-Ray Densitometers for Bone Mineral Measurements.

Authors:  J Saarelainen; M Hakulinen; T Rikkonen; H Kröger; M Tuppurainen; H Koivumaa-Honkanen; R Honkanen; M Hujo; J S Jurvelin
Journal:  J Osteoporos       Date:  2016-04-27

2.  Comparison of screening strategies for prevalent vertebral fractures in South Korea: vertebral fracture assessment vs. spine radiography.

Authors:  Sung-Hee Oh; Dam Kim; Young Eun Lee; Deog-Yoon Kim; Yu Kyung Lee; Joo-Hyun Lee; Sang-Cheol Bae; Yun Young Choi; Junhee Pyo; Jeonghoon Ahn; Yoon-Kyoung Sung
Journal:  BMC Musculoskelet Disord       Date:  2018-02-12       Impact factor: 2.362

3.  Does Bone Mineral Density Differ between Fan-Beam and Pencil-Beam?: A Meta-Analysis and Systemic Review.

Authors:  Byung-Ho Yoon; Young Do Koh; Jun-Il Yoo; Sujin Kim; Guen Young Lee; Sung Bin Park; Yong-Chan Ha
Journal:  J Bone Metab       Date:  2021-02-28

4.  Cross-calibration of two dual-energy X-ray absorptiometry devices for the measurement of body composition in young children.

Authors:  Jaz Lyons-Reid; Timothy Kenealy; Benjamin B Albert; Kate A Ward; Nicholas Harvey; Keith M Godfrey; Shiao-Yng Chan; Wayne S Cutfield
Journal:  Sci Rep       Date:  2022-08-16       Impact factor: 4.996

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.