Comparison of a new portable digital meniscometer and optical coherence tomography in tear meniscus radius measurement.

Stefan Bandlitz1, Christine Purslow, Paul J Murphy, Heiko Pult.   

Abstract

PURPOSE: Non-invasive measurement of tear meniscus radius (TMR) is useful in the assessment of tear volume for dry eye diagnosis. This study investigates the agreement between a new, portable, slit-lamp mounted, digital meniscometer (PDM) and optical coherence tomography (OCT) in the measurement of human TMR.
METHODS: Images of the tear meniscus at the centre of the lower lid of 30 normal subjects (8M, 22F; mean age 27.5 SD ± 9.6 years) were taken using the PDM and the OCT. On the PDM and OCT images, TMR was measured using imagej 1.46b software. The meniscus on the OCT images was subdivided vertically into three equal sections and the radius calculated for each: bottom (BTMR), centre (CTMR) and top (TTMR). The relationship between PDM and OCT measurements was analysed using Spearman's rank coefficient, and differences between PDM and OCT subsection measurements were evaluated using Bland-Altman plots.
RESULTS: Tear meniscus radius measured with the PDM (0.25 ± 0.06 mm) and OCT (0.29 ± 0.09 mm) was significantly correlated (r = 0.675; p < 0.001). The mean differences between TMR using the PDM and the subsections from OCT showed that TMR measured with PDM was greater for BTMR (0.07 mm; CI 0.05-0.10; p < 0.001), similar for CTMR (-0.01 mm; CI -0.04 to 0.02; p = 0.636) and steeper for TTMR (-0.07 mm; CI -0.10 to -0.04; p < 0.001).
CONCLUSIONS: Portable digital meniscometer and OCT measurements of the TMR are significantly correlated, suggesting that the new PDM is a useful surrogate for OCT in this respect. The PDM appears to measure the radius of the central section of the tear meniscus.
© 2013 Acta Ophthalmologica Scandinavica Foundation. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  optical coherence tomography; portable digital meniscometer; reflective meniscometry; tear meniscus radius; tear volume

Mesh:

Year:  2013        PMID: 24119097     DOI: 10.1111/aos.12275

Source DB:  PubMed          Journal:  Acta Ophthalmol        ISSN: 1755-375X            Impact factor:   3.761


  3 in total

1.  The relationship of tear osmolarity with tear meniscus curvature and contact angles in healthy subjects: anterior segment optical coherence tomography study.

Authors:  Volkan Yeter; Nurullah Koçak; Merve Kalyoncu; Ertuğrul Can; Nurşen Arıtürk
Journal:  Int Ophthalmol       Date:  2021-09-29       Impact factor: 2.031

Review 2.  Diagnosis of dry eye disease and emerging technologies.

Authors:  Maya Salomon-Ben Zeev; Darby Douglas Miller; Robert Latkany
Journal:  Clin Ophthalmol       Date:  2014-03-20

3.  Tear volume estimation using a modified Schirmer test: a randomized, multicenter, double-blind trial comparing 3% diquafosol ophthalmic solution and artificial tears in dry eye patients.

Authors:  Hideki Miyake; Yuri Kawano; Hiroshi Tanaka; Akihiro Iwata; Takahiro Imanaka; Masatsugu Nakamura
Journal:  Clin Ophthalmol       Date:  2016-05-13
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.