UK multicenter experience using the Gore septal occluder (GSO(TM) ) for atrial septal defect closure in children and adults.

Ben Smith1, John Thomson, David Crossland, Mark S Spence, Gareth J Morgan.   

Abstract

BACKGROUND: Percutaneous closure of atrial defects (ASD) has evolved as the treatment of choice for the majority of defects and patent oval foramens. The Gore Septal Occluder (GSO) is an innovative device consisting mostly of a folded thin GoreTex tube for use in the closure of septal defects.
METHODS: Reviewed is the multicenter UK experience of the first 22 ASD occlusions with the GSO(TM) device. All implantations were performed by consultant operators experienced in ASD device closure. The inclusion criterion was the presence of a hemodynamically significant secundum ASD with a diameter of <18 mm. Procedural data and acute and mid-term closure rates were retrospectively matched to a cohort of patients having defect closure using the Amplatzer Septal Occluder(TM) (ASO(TM) ).
RESULTS: Acute and 3-month follow-up closure rates for the GSO(TM) were 100% and 100% vs. 100% and 100% closure with the ASO(TM) implants. The difference in paired procedure times was not statistically significant (56 min: GSO(TM) ; 42 min: ASO(TM) device P = ns), nor was the paired difference in fluoroscopic screening times (12 min: GSO(TM) vs. 8.4 min: ASO(TM) , P = ns). One GSO device embolized immediately after deployment and was successfully retrieved at the same procedure. There were no other significant complications in either group.
CONCLUSIONS: The GSO(TM) implant can achieve comparable closure rates to the ASO(TM) in small to moderate atrial septal defects after 3 months. Longer fluoroscopy and procedure times are a drawback; however these should improve with familiarity with the implant and deployment system. The larger sheath size was not associated with increased complications in our cohort.
Copyright © 2013 Wiley Periodicals, Inc.

Entities:  

Keywords:  atrial septal defect; congenital interventional cardiology; device occlusion

Mesh:

Substances:

Year:  2013        PMID: 24115686     DOI: 10.1002/ccd.25216

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  4 in total

1.  Transcatheter Closure of Atrial Septal Defects using the GORE(®) Septal Occluder in Children Less Than 10 kg of Body Weight.

Authors:  Tariq Abu-Tair; Christiane M Wiethoff; Jascha Kehr; Wlodzimierz Kuroczynski; Christoph Kampmann
Journal:  Pediatr Cardiol       Date:  2016-02-19       Impact factor: 1.655

Review 2.  Transcatheter closure of atrial septal defect: principles and available devices.

Authors:  Se Yong Jung; Jae Young Choi
Journal:  J Thorac Dis       Date:  2018-09       Impact factor: 2.895

3.  Late complete atrioventricular block after closure of an atrial septal defect with a gore septal occluder (GSO™).

Authors:  Sven Dittrich; Matthias Sigler; Helga Priessmann
Journal:  Catheter Cardiovasc Interv       Date:  2015-09-10       Impact factor: 2.692

Review 4.  Translation in cardiovascular stents and occluders: From biostable to fully degradable.

Authors:  Yingying Huang; Yee Shan Wong; Herr Cheun Anthony Ng; Freddy Y C Boey; Subbu Venkatraman
Journal:  Bioeng Transl Med       Date:  2017-07-19
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.