Literature DB >> 24111039

Emergency Life Support System aiming preprimed oxygenator.

Takashi Isoyama, Koki Ariyoshi, Kyosuke Nii, Itsuro Saito, Kazuyoshi Fukunaga, Yusuke Inoue, Toshiya Ono, Kohei Ishii, Shintaro Hara, Kou Imachi, Madoka Takai, Yusuke Abe.   

Abstract

Development have been achieved of a new blood pump for next generation Percutaneous Cardio-Pulmonary Support (PCPS) system and a novel surface coating method for silicone membrane hollow fiber by physical adsorption using a copolymer composed of a 2-Methacryloyloxyethyl phosphorylcholine (MPC) unit and a hydrophobic unit. The new blood pump, named the Troidal Convolution Pump (TCP), is based on the principle of a cascade pump and perfused 5 L/min and 350 mmHg at 2450 rpm. The novel copolymer composed of 30% MPC unit and 3-(methacryloyloxy) propyltris (trimethylsiloxy) silane (MPTSSi) unit (PMMSi30) was the most suitable molecular design on a silicone surface. The PMMSi30 coated surface adsorbed 7.2 % as much protein a non-coated surface adsorbed.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24111039     DOI: 10.1109/EMBC.2013.6610852

Source DB:  PubMed          Journal:  Conf Proc IEEE Eng Med Biol Soc        ISSN: 1557-170X


  1 in total

1.  Principle and basic property of the sequential flow pump.

Authors:  Shintaro Hara; Erina Maeno; Xinyang Li; Terumi Yurimoto; Takashi Isoyama; Itsuro Saito; Toshiya Ono; Yusuke Abe
Journal:  J Artif Organs       Date:  2017-04-19       Impact factor: 1.731