Characterization of small-field stereotactic radiosurgery beams with modern detectors.

Madelaine Tyler1, Paul Z Y Liu, Kin Wa Chan, Anna Ralston, David R McKenzie, Simon Downes, Natalka Suchowerska.   

Abstract

To derive accurate beam models for stereotactic radiosurgery (SRS) planning it is necessary to characterize the beam with dosimetric measurements. The aim of this study is to identify the best detectors for each task in the characterization process. Output ratios, beam profiles and percentage depth doses were measured for SRS cone diameters of 5-45 mm. Commercially available and emerging detectors were used: Gafchromic EBT2 film, an air-core fibre optic dosimeter (FOD) (developed at Royal Prince Alfred Hospital, Sydney), an IBA stereotactic field diode, a PTW 60012 electron diode and an IBA cc01 small volume thimble ion chamber. Analysis of the measured data supported by baseline Monte Carlo simulation data, led to the following recommendations: (1) water-equivalent detectors (Gafchromic EBT2 film or FOD) are the preferred choice for SRS dosimetry, (2) ion chambers (including small volume chambers with high-density central electrodes) should be avoided due to volume averaging effects and energy dependence, (3) if diodes are used, corrections must be made to account for their over-response in small fields.

Entities:  

Mesh:

Year:  2013        PMID: 24107592     DOI: 10.1088/0031-9155/58/21/7595

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  11 in total

1.  Analysis of small field percent depth dose and profiles: Comparison of measurements with various detectors and effects of detector orientation with different jaw settings.

Authors:  Henry Finlay Godson; M Ravikumar; S Sathiyan; K M Ganesh; Y Retna Ponmalar; C Varatharaj
Journal:  J Med Phys       Date:  2016 Jan-Mar

2.  Deriving detector-specific correction factors for rectangular small fields using a scintillator detector.

Authors:  Yujiao Qin; Hualiang Zhong; Ning Wen; Karen Snyder; Yimei Huang; Indrin J Chetty
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

3.  Small field detector correction factors: effects of the flattening filter for Elekta and Varian linear accelerators.

Authors:  Madelaine K Tyler; Paul Z Y Liu; Christopher Lee; David R McKenzie; Natalka Suchowerska
Journal:  J Appl Clin Med Phys       Date:  2016-05-08       Impact factor: 2.102

4.  Detectors assessment for stereotactic radiosurgery with cones.

Authors:  Nicolas Garnier; Régis Amblard; Rémy Villeneuve; Rodolphe Haykal; Cécile Ortholan; Philippe Colin; Anaïs Gérard; Sarah Belhomme; Franck Mady; Mourad Benabdesselam; Benjamin Serrano
Journal:  J Appl Clin Med Phys       Date:  2018-09-14       Impact factor: 2.102

5.  "Characterization of ELEKTA SRS cone collimator using high spatial resolution monolithic silicon detector array".

Authors:  Khalsa Al Shukaili; Stéphanie Corde; Marco Petasecca; Vladimir Pereveratylo; Michael Lerch; Michael Jackson; Anatoly Rosenfeld
Journal:  J Appl Clin Med Phys       Date:  2018-05-22       Impact factor: 2.102

6.  Real-time high spatial resolution dose verification in stereotactic motion adaptive arc radiotherapy.

Authors:  Mitchell Duncan; Matthew K Newall; Vincent Caillet; Jeremy T Booth; Paul J Keall; Michael Lerch; Vladimir Perevertaylo; Anatoly B Rosenfeld; Marco Petasecca
Journal:  J Appl Clin Med Phys       Date:  2018-06-05       Impact factor: 2.102

7.  Evaluation of the IAEA-TRS 483 protocol for the dosimetry of small fields (square and stereotactic cones) using multiple detectors.

Authors:  Clare L Smith; Atousa Montesari; Christopher P Oliver; Duncan J Butler
Journal:  J Appl Clin Med Phys       Date:  2019-12-30       Impact factor: 2.102

8.  Assessment of Delivery Quality Assurance for Stereotactic Radiosurgery With Cyberknife.

Authors:  Jun Li; Xile Zhang; Yuxi Pan; Hongqing Zhuang; Junjie Wang; Ruijie Yang
Journal:  Front Oncol       Date:  2021-11-17       Impact factor: 6.244

9.  CyberKnife® fixed cone and Iris™ defined small radiation fields: Assessment with a high-resolution solid-state detector array.

Authors:  Giordano Biasi; Marco Petasecca; Susanna Guatelli; Ebert A Martin; Garry Grogan; Benjamin Hug; Jonathan Lane; Vladimir Perevertaylo; Tomas Kron; Anatoly B Rosenfeld
Journal:  J Appl Clin Med Phys       Date:  2018-07-12       Impact factor: 2.102

10.  Feasibility study of using Stereotactic Field Diode for field output factors measurement and evaluating three new detectors for small field relative dosimetry of 6 and 10 MV photon beams.

Authors:  Attia Gul; Shigekazu Fukuda; Hideyuki Mizuno; Nakaji Taku; M Basim Kakakhel; Sikander M Mirza
Journal:  J Appl Clin Med Phys       Date:  2020-10-19       Impact factor: 2.243

View more