Literature DB >> 24107326

Mechanism of edge restenosis after drug-eluting stent implantation. Angulation at the edge and mechanical properties of the stent.

Yun Gi Kim1, Il-Young Oh, Yoo-Wook Kwon, Jung-Kyu Han, Han-Mo Yang, Kyung-Woo Park, Hae-Young Lee, Hyun-Jae Kang, Bon-Kwon Koo, Hyo-Soo Kim.   

Abstract

BACKGROUND: Edge restenosis is not an unusual finding after implantation of drug-eluting stents (DES). We hypothesized that mechanical stress imposed on the stent edge would cause vessel wall injury and inflammation, which may consequently lead to edge restenosis. METHODS AND
RESULTS: In total, 1,496 patients were implanted with a sirolimus-eluting stent (SES), paclitaxel-eluting stent (PES) or zotarolimus-eluting stent (ZES) in Seoul National University Hospital between 2007 and 2009. Binary restenosis occurred in 161 lesions in 119 patients. We retrospectively compared the 3 DES with regard to the percentage of edge stent restenosis among all cases of restenosis. We also evaluated the maximal, minimal, and Δ (maximal angle-minimal angle) angles. The percentage of edge restenosis was higher for SES than for ZES (37.5% vs. 16.7%, P=0.017). Maximal angle at the proximal edge was 64.82°±33.46° for 26 stents with proximal edge restenosis compared with 31.84°±31.51° for 89 stents without proximal edge restenosis (P=0.001). The Δ angle was also significantly different between the 2 groups (14.81°±15.98° vs. 7.60°±8.86°, P=0.035). Similar findings were observed for distal edge restenosis. Both the maximal angle (39.09°±21.04° vs. 22.71°±22.83°, P=0.010) and Δ angle (20.23°±15.39° vs. 9.18°±9.66°, P=0.016) at the distal edge were significantly different between the 2 groups.
CONCLUSIONS: Physical stress determined by angulation at the stent edge segment and biomechanical properties of the DES can be considered as one of the plausible mechanisms for edge stent restenosis.

Entities:  

Mesh:

Year:  2013        PMID: 24107326     DOI: 10.1253/circj.cj-12-1259

Source DB:  PubMed          Journal:  Circ J        ISSN: 1346-9843            Impact factor:   2.993


  3 in total

1.  Higher plasma level of STIM1, OPG are correlated with stent restenosis after PCI.

Authors:  Haibin Li; Zhian Jiang; Xiangdong Liu; Zhihui Yang
Journal:  Int J Clin Exp Med       Date:  2015-11-15

2.  Stentless Strategy by Drug-Coated Balloon Angioplasty following Directional Coronary Atherectomy for Left Main Bifurcation Lesion.

Authors:  Norihiro Kobayashi; Masahiro Yamawaki; Shinsuke Mori; Masakazu Tsutsumi; Yohsuke Honda; Kenji Makino; Shigemitsu Shirai; Masafumi Mizusawa; Yoshiaki Ito
Journal:  J Interv Cardiol       Date:  2021-03-03       Impact factor: 2.279

3.  Influence of Stent Flexibility on Artery Wall Stress and Wall Shear Stress in Bifurcation Lesions.

Authors:  Noboru Saito; Yuhei Mori; Tomoya Komatsu
Journal:  Med Devices (Auckl)       Date:  2020-11-02
  3 in total