| Literature DB >> 24103352 |
Andrej Teren1, Frank Beutner, Kerstin Wirkner, Markus Loeffler, Markus Scholz.
Abstract
BACKGROUND: Ankle-brachial-Index (ABI) measured by manual Dopplersonography is an easily assessable marker of global cardiovascular risk. The aim of this study was to establish novel photo-plethysmography (PPG)-based ABI assessments in an epidemiologic context and to compare its results with those of Doppler.Entities:
Mesh:
Year: 2013 PMID: 24103352 PMCID: PMC3851735 DOI: 10.1186/1471-2261-13-81
Source DB: PubMed Journal: BMC Cardiovasc Disord ISSN: 1471-2261 Impact factor: 2.298
Figure 1Illustration of ABI determination by Vicorder and Vascular Explorer. We present the default methods of both PPG-based devices. While Vascular Explorer measures during the disappearance of signal (inflation method), Vicorder measures during the re-appearance of the signal (deflation method).
Study design
| 1 | D-A | VE - A | VI - A | VE – B | VI – B | 7 |
| 2 | D-A | VI - A | VE - A | VI – B | VE – B | 7 |
| 3 | D-B | VE – B | VI - B | VE - A | VI – A | 7 |
| 4 | D-B | VI – B | VE – B | VI - A | VE – A | 7 |
| 5 | D-A | VE - A | VI - A | VE - A | VI – A | 7 |
| 6 | D-A | VI - A | VE - A | VI - A | VE – A | 7 |
| 7 | D-B | VE - B | VI - B | VE - B | VI – B | 7 |
| 8 | D-B | VI - B | VE - B | VI - B | VE – B | 7 |
Assessments were performed in the prescribed consecutive order. D = Doppler sonography, VI = Vicorder, VE = Vascular explorer, A,B = different observers.
Figure 2Agreement of automated ABI measurements those of Doppler sonography. We present the raw data of both devices. Both automated ABI measurements are clearly biased towards higher values which is more pronounced for Vascular Explorer.
Intra-observer concordances of ABI measurements
| VI | 0.65 | 0.46 | 0.78 | 0.015 | 0.29 | 0.65 |
| VI_O | 0.68 | 0.51 | 0.80 | 0.001 | 0.95 | 0.68 |
| VI_E | 0.65 | 0.47 | 0.78 | 0.006 | 0.65 | 0.66 |
| VE | 0.65 | 0.50 | 0.76 | -0.011 | 0.43 | 0.66 |
| VE_O | 0.75 | 0.59 | 0.85 | -0.004 | 0.75 | 0.75 |
| VE_E | 0.76 | 0.57 | 0.87 | -0.002 | 0.89 | 0.76 |
| VE_D | 0.57 | 0.27 | 0.78 | 0.021 | 0.12 | 0.59 |
Agreement of first and second ABI measurement for automated methods. We present concordance correlation coefficients, corresponding confidence intervals, bias between first and second measurements (values less than zero mean lower values at first measurement), p-value of bias and correlation coefficient of first and second measurement.
VI/VE = Vicorder/Vascular Explorer standard measurements, VI_O/VE_O = Vicorder/Vascular Explorer corrected by observer, VI_E/VE_E = Vicorder/Vascular Explorer corrected by examiner, VE_D = deflation method of Vascular Explorer.
Figure 3Bland-Altman plots for intra-observer agreement. We present the results of the raw ABI measurements for both Vicorder and Vascular Explorer.
Inter-observer concordances of ABI measurements
| VI | 0.70 | 0.50 | 0.83 | -0.019 | 0.18 | 0.71 |
| VI_O | 0.71 | 0.51 | 0.84 | -0.032 | 0.017 | 0.73 |
| VI_E | 0.73 | 0.53 | 0.85 | -0.026 | 0.046 | 0.74 |
| VE | 0.55 | 0.31 | 0.73 | -0.00042 | 0.98 | 0.55 |
| VE_O | 0.49 | 0.23 | 0.69 | -0.010 | 0.46 | 0.50 |
| VE_E | 0.54 | 0.21 | 0.76 | -0.024 | 0.060 | 0.55 |
| VE_D | 0.56 | 0.16 | 0.89 | -0.023 | 0.16 | 0.58 |
Agreement of first and second ABI measurement for automated methods. We present concordance correlation coefficients, corresponding confidence intervals, bias between observers, p-value of bias and correlation coefficient between observers.
VI/VE = Vicorder/Vascular Explorer standard measurements, VI_O/VE_O = Vicorder/Vascular Explorer corrected by observer, VI_E/VE_E = Vicorder/Vascular Explorer corrected by examiner, VE_D = Deflation method of Vascular Explorer.
Figure 4Bland-Altman plots for inter-observer agreement. We present the results of the raw ABI measurements for both Vicorder and Vascular Explorer.