Literature DB >> 24102186

Low-tilt monophasic and biphasic waveforms compared with standard biphasic waveforms in the transvenous defibrillation of ventricular fibrillation.

Johan R Bennett1, Karen M Darragh, Simon J Walsh, Desmond J Allen, Michael Scott, Michael Stevenson, Jennifer A A Adgey, John M C J Anderson, Ganesh Manoharan.   

Abstract

BACKGROUND: Commercially available implantable defibrillators utilize a high-tilt waveform. Studies in atrial fibrillation and transthoracic defibrillation of ventricular fibrillation (VF) have shown improved defibrillation efficacy using low-tilt (LT) waveforms. We investigated the feasibility, efficacy, and safety of a LT waveform in the transvenous defibrillation of VF and hypothesized that it would be more efficacious than standard tilted biphasic (STB) waveforms. METHODS AND
RESULTS: The investigation was performed in four phases in a porcine model: an efficacy study of LT monophasic waveforms (n = 9), an efficacy study of LT biphasic waveforms (n = 9), a comparison study between the most successful LT waveforms and clinically available STB waveforms (n = 15), and a safety study (n = 9). A total of 1,056 shocks were delivered (phase 1: 288, phase 2: 288, phase 3: 480). The LT biphasic 8/4-ms waveform was significantly more likely to successfully defibrillate than the LT monophasic and STB waveforms with an odds ratio of 122.3 (95% confidence interval: 32.5, 460.2, P < 0.001). The calculated defibrillation threshold (E50) for the LT 8/4-ms waveform was 12.7 J compared to 43.5 J and 45.5 J for STB waveforms 6/6 ms and 8/4 ms, respectively, and 47.7 J for LT 12-ms waveform. The LT 8/4-ms waveform had no lasting detrimental effect on cardiac function, and any transient hemodynamical or biochemical changes observed were comparable to those observed with STB waveforms.
CONCLUSION: LT waveforms are effective and appear safe in transvenous defibrillation in a porcine model of VF. The LT biphasic 8/4-ms waveform is more efficacious than conventional waveforms. ©2013, The Authors. Journal compilation ©2013 Wiley Periodicals, Inc.

Entities:  

Keywords:  defibrillation; low tilt; ventricular fibrillation; waveform

Mesh:

Year:  2013        PMID: 24102186     DOI: 10.1111/pace.12266

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  1 in total

1.  Relationship Between Heart Rate Variability and Pulse Rate Variability Measures in Patients After Coronary Artery Bypass Graft Surgery.

Authors:  Yung-Sheng Chen; Yi-Ying Lin; Chun-Che Shih; Cheng-Deng Kuo
Journal:  Front Cardiovasc Med       Date:  2021-12-16
  1 in total