Reduction of effective dose and organ dose to the eye lens in head MDCT using iterative image reconstruction and automatic tube current modulation.

Pavel Ryska1, Tomas Kvasnicka, Jiri Jandura, Ludovit Klzo, Jakub Grepl, Jan Zizka.   

Abstract

AIMS: To compare the effective and eye lens radiation dose in helical MDCT brain examinations using automatic tube current modulation in conjunction with either standard filtered back projection (FBP) technique or iterative reconstruction in image space (IRIS).
METHODS: Of 400 adult brain MDCT examinations, 200 were performed using FBP and 200 using IRIS with the following parameters: tube voltage 120 kV, rotation period 1 second, pitch factor 0.55, automatic tube current modulation in both transverse and longitudinal planes with reference mAs 300 (FBP) and 200 (IRIS). Doses were calculated from CT dose index and dose length product values utilising ImPACT software; the organ dose to the lens was derived from the actual tube current-time product value applied to the lens. Image quality was assessed by two independent readers blinded to the type of image reconstruction technique.
RESULTS: The average effective scan dose was 1.47±0.26 mSv (FBP) and 0.98±0.15 mSv (IRIS), respectively (33.3% decrease). The average organ dose to the eye lens decreased from 40.0±3.3 mGy (FBP) to 26.6±2.0 mGy (IRIS, 33.5% decrease). No significant change in diagnostic image quality was noted between IRIS and FBP scans (P=0.17).
CONCLUSION: Iterative reconstruction of cerebral MDCT examinations enables reduction of both effective and organ eye lens dose by one third without signficant loss of image quality.

Mesh:

Year:  2013        PMID: 24091879     DOI: 10.5507/bp.2013.071

Source DB:  PubMed          Journal:  Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub        ISSN: 1213-8118            Impact factor:   1.245


  1 in total

1.  Lens Dose Reduction by Patient Posture Modification During Neck CT.

Authors:  Elizabeth G Mosher; John A Butman; Les R Folio; Nadia M Biassou; Choonsik Lee
Journal:  AJR Am J Roentgenol       Date:  2018-03-16       Impact factor: 3.959

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.