| Literature DB >> 24089863 |
A Bodin1, R Laloo, P Abeilhou, L Guiraud, S Gauthier, D Martrou.
Abstract
We have developed an energy-filtering device coupled to a quadrupole mass spectrometer to deposit ionized molecules on surfaces with controlled energy in ultra high vacuum environment. Extensive numerical simulations as well as direct measurements show that the ion beam flying out of a quadrupole exhibits a high-energy tail decreasing slowly up to several hundred eV. This energy distribution renders impossible any direct soft-landing deposition of molecular ions. To remove this high-energy tail by energy filtering, a 127° electrostatic sector and a specific triplet lenses were designed and added after the last quadrupole of a triple quadrupole mass spectrometer. The results obtained with this energy-filtering device show clearly the elimination of the high-energy tail. The ion beam that impinges on the sample surface satisfies now the soft-landing criterion for molecular ions, opening new research opportunities in the numerous scientific domains involving charges adsorbed on insulating surfaces.Entities:
Year: 2013 PMID: 24089863 DOI: 10.1063/1.4818961
Source DB: PubMed Journal: Rev Sci Instrum ISSN: 0034-6748 Impact factor: 1.523