Literature DB >> 24088060

Assessment of axial length before and after myopic LASIK with the IOLMaster.

Eugene Tay, Xiang Li, Howard V Gimbel, Geoffrey Kaye.   

Abstract

PURPOSE: To assess the relationship between theoretical ablation depth and axial length change after LASIK with the IOLMaster (Carl Zeiss Meditec, Dublin, CA).
METHODS: Ninety-nine eyes were examined preoperatively and 1 and 3 months after LASIK.
RESULTS: Mean ± standard deviation (SD) spherical equivalent before LASIK was -4.06 ± 1.91 diopters (D). Mean ± SD ablation depth was 83.13 ± 30.31 μm. Mean ± SD postoperative axial length of 25.11 ± 0.14 mm at 1 month was significantly shorter than mean ± SD preoperative axial length of 25.20 ± 0.14 mm (P < .001) with no subsequent change thereafter (P = .450). An increase in ablation depth of 1 μm led to a decrease in axial length of 0.00118 ± 0.00005 mm. Ablation depth correlated strongly with change in axial length (adjusted R(2) = 0.9039).
CONCLUSIONS: The IOLMaster showed a decrease in axial length after LASIK that correlated well with theoretical ablation depth. Copyright 2013, SLACK Incorporated.

Mesh:

Year:  2013        PMID: 24088060     DOI: 10.3928/1081597X-20130924-01

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  3 in total

1.  Measurement of crystalline lens tilt in high myopic eyes before cataract surgery using swept-source optical coherence tomography.

Authors:  Qiang Lu; Wenwen He; Dongjin Qian; Yi Lu; Xiangjia Zhu
Journal:  Eye Vis (Lond)       Date:  2020-03-06

2.  Long-Term Visual and Refractive Stability and Ocular Biometric Changes after Laser-Assisted Subepithelial Keratomileusis for Correction of Myopia: An 8-Year Follow-Up.

Authors:  Seyed Javad Hashemian; Acieh Es'haghi; Parya Abdolalizadeh; Leila Ghiasian; Hossein Aghaei; Mohammad Ebrahim Jafari; Mahsa Sadat Hashemian; Seyed Mahyar Hashemian
Journal:  J Curr Ophthalmol       Date:  2022-01-06

3.  Precision (repeatability and reproducibility) of ocular parameters obtained by the Tomey OA-2000 biometer compared to the IOLMaster in healthy eyes.

Authors:  Yanjun Hua; Wei Qiu; Qiuyi Xiao; Qiang Wu
Journal:  PLoS One       Date:  2018-02-27       Impact factor: 3.240

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.