Literature DB >> 24076257

Total monitor units influence on plan quality parameters in volumetric modulated arc therapy for breast case.

P Mancosu1, G Reggiori2, F Alongi2, L Cozzi3, A Fogliata3, F Lobefalo2, P Navarria2, A Stravato2, S Tomatis2, M Scorsetti2.   

Abstract

PURPOSE: To investigate the correlation between total monitor units (MU), dosimetric findings, and pre-treatment quality assurance for volumetric modulated arc therapy (VMAT) by RapidArc (RA). METHODS AND MATERIALS: Ten patients with breast cancer were considered. Dose prescriptions were: 48 Gy and 40.5 Gy in 15 fractions to, respectively, PTV(Boost) and PTVWholeBreast. A reference plan was optimized and four more plans using the "MU Objective", a tool for total MU controlling, were prepared imposing ± 20 and ± 50% total MU for inducing different complexities. Plan objectives were: D95% > 95% for both PTVs, and D2% < 107% for PTV(Boost); mean dose < 9.5 Gy and V20 Gy < 10% for ipsilateral lung; V18 Gy < 5% for heart; mean dose <3 Gy for controlateral breast; furthermore V5 Gy, V10 Gy, V20 Gy, and V30 Gy to body were minimized. Plans were evaluated in terms of technical parameters, dosimetric plan objectives findings and pre-treatment quality assurance (QA).
RESULTS: Concerning PTVs, there were no significant differences for target coverage (D95%); mean doses for ipsilateral lung and controlateral breast, and V18 Gy for heart decreased with MUs increasing, reaching a plateau with reference plan. Body volume receiving low dose (V5-10 Gy) was minimized for reference plans. All plans had GAI (3 mm, 3%) > 95%.
CONCLUSIONS: The data suggest that the best plan is the reference one, where the "MU Objective" tool was not used during optimisation. Nevertheless, it is advisable to use the "MU Objective" tool for re-planning when low GAI is found to increase its value. In this case, attention should be paid to OARs dose limits, since their values may be increased.
Copyright © 2013 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Plan quality; Radiotherapy; RapidArc; Volumetric modulated arc therapy

Mesh:

Year:  2013        PMID: 24076257     DOI: 10.1016/j.ejmp.2013.08.142

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  4 in total

1.  Performance of the eclipse monitor unit objective tool utilizing volumetric modulated arc therapy for rectal cancer.

Authors:  Alejandro Prado; Ángel Gaitán; Mario Leonor; Marta Manzano; Eduardo Cabello; Raúl Díaz; Alejandro Ferrando; Ana Milanés; Gustavo Pozo
Journal:  Rep Pract Oncol Radiother       Date:  2019-02-25

Review 2.  Critical appraisal of the role of volumetric modulated arc therapy in the radiation therapy management of breast cancer.

Authors:  Luca Cozzi; Frank Lohr; Antonella Fogliata; Davide Franceschini; Fiorenza De Rose; A R Filippi; Gabriele Guidi; Valentina Vanoni; Marta Scorsetti
Journal:  Radiat Oncol       Date:  2017-12-19       Impact factor: 3.481

3.  Limiting treatment plan complexity by applying a novel commercial tool.

Authors:  Alessandro Scaggion; Marco Fusella; Giancarmelo Agnello; Andrea Bettinelli; Nicola Pivato; Antonella Roggio; Marco A Rossato; Matteo Sepulcri; Marta Paiusco
Journal:  J Appl Clin Med Phys       Date:  2020-05-21       Impact factor: 2.102

4.  The use of collimator angle optimization and jaw tracking for VMAT-based single-isocenter multiple-target stereotactic radiosurgery for up to six targets in the Varian Eclipse treatment planning system.

Authors:  Lauren M M Pudsey; Dean Cutajar; Alex Wallace; Anastasia Saba; Laurel Schmidt; Andrej Bece; Catherine Clark; Yoshiya Yamada; Giordano Biasi; Anatoly Rosenfeld; Joel Poder
Journal:  J Appl Clin Med Phys       Date:  2021-07-19       Impact factor: 2.102

  4 in total