Marker-based validation of a biplane fluoroscopy system for quantifying foot kinematics.

Joseph M Iaquinto1, Richard Tsai2, David R Haynor3, Michael J Fassbind2, Bruce J Sangeorzan4, William R Ledoux5.   

Abstract

INTRODUCTION: Radiostereometric analysis has demonstrated its capacity to track precise motion of the bones within a subject during motion. Existing devices for imaging the body in two planes are often custom built systems; we present here the design and marker-based validation of a system that has been optimized to image the foot during gait.
METHODS: Mechanical modifications were made to paired BV Pulsera C-arms (Philips Medical Systems) to allow unfettered gait through the imaging area. Image quality improvements were obtained with high speed cameras and the correction of image distorting artifacts. To assess the system's accuracy, we placed beads at known locations throughout the imaging field, and used post processing software to calculate their apparent locations.
RESULTS: Distortion correction reduced overall RMS error from 6.56 mm to 0.17 mm. When tracking beads in static images a translational accuracy of 0.094 ± 0.081 mm and rotational accuracy of 0.083 ± 0.068° was determined. In dynamic trials simulating speeds seen during walking, accuracy was 0.126 ± 0.122 mm. DISCUSSION: The accuracies and precisions found are within the reported ranges from other such systems. With the completion of marker-based validation, we look to model-based validation of the foot during gait. Published by Elsevier Ltd.

Keywords:  Biomechanics; Biplane fluoroscopy; Foot; X-ray

Mesh:

Year:  2013        PMID: 24075068     DOI: 10.1016/j.medengphy.2013.08.013

Source DB:  PubMed          Journal:  Med Eng Phys        ISSN: 1350-4533            Impact factor:   2.242


  4 in total

1.  Error performances of a model-based biplane fluoroscopic system for tracking knee prosthesis during treadmill gait task.

Authors:  Arnaud Barré; Kamiar Aminian
Journal:  Med Biol Eng Comput       Date:  2017-07-18       Impact factor: 2.602

2.  The Glasgow-Maastricht foot model, evaluation of a 26 segment kinematic model of the foot.

Authors:  Michiel Oosterwaal; Sylvain Carbes; Scott Telfer; James Woodburn; Søren Tørholm; Amir A Al-Munajjed; Lodewijk van Rhijn; Kenneth Meijer
Journal:  J Foot Ankle Res       Date:  2016-07-08       Impact factor: 2.303

3.  Measuring 3D Hand and Finger Kinematics-A Comparison between Inertial Sensing and an Opto-Electronic Marker System.

Authors:  Josien C van den Noort; Henk G Kortier; Nathalie van Beek; DirkJan H E J Veeger; Peter H Veltink
Journal:  PLoS One       Date:  2016-11-03       Impact factor: 3.240

4.  A Direct Comparison of Biplanar Videoradiography and Optical Motion Capture for Foot and Ankle Kinematics.

Authors:  Sarah E Kessler; Michael J Rainbow; Glen A Lichtwark; Andrew G Cresswell; Susan E D'Andrea; Nicolai Konow; Luke A Kelly
Journal:  Front Bioeng Biotechnol       Date:  2019-08-23
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.