Inter- and intraobserver variability in resonance frequency analysis of palatal implants--a technical note.

Michael H Bertl, Ruüdiger Emshoff, Aleš Čelar, Adriano G Crismani.   

Abstract

PURPOSE: To assess inter- and intraobserver variability in wireless resonance frequency analysis (RFA) stability measurements of palatal implants and to evaluate the influence of age, sex, time after implant insertion, and measurement direction on variability.
MATERIALS AND METHODS: Three observers conducted wireless RFA stability measurements of palatal implants from 16 patients. Measurements were taken in anteroposterior and laterolateral directions and were repeated after 1 hour.
RESULTS: Data showed a small interobserver variation (1.93) with intraobserver variation (1.77) as its largest component. Time after implantation showed a strong influence (P = .027) on the interobserver variation.
CONCLUSIONS: The tested wireless RFA device can be considered a reliable instrument for measuring the stability of palatal implants. Multiple readings and observers could further improve measurement reliability.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24066338     DOI: 10.11607/jomi.2831

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804