Ikuma Fukuda1, Hiroaki Hayashi, Kazuki Takegami, Yuki Konishi.
Abstract
Diagnostic X-ray equipment was used to develop an experimental apparatus for calibrating a CdTe detector. Powder-type samples were irradiated with collimated X-rays. On excitation of the atoms, characteristic X-rays were emitted. We prepared Nb2O5, SnO2, La2O3, Gd2O3, and WO3 metal oxide samples. Experiments using the diagnostic X-ray equipment were carried out to verify the practicality of our apparatus. First, we verified that the collimators involving the apparatus worked well. Second, the X-ray spectra were measured using the prepared samples. Finally, we analyzed the spectra, which indicated that the energy calibration curve had been obtained at an accuracy of ±0.06 keV. The developed apparatus could be used conveniently, suggesting it to be useful for the practical training of beginners and researchers.Entities:
Mesh:
Substances:
Year: 2013 PMID: 24064699 DOI: 10.6009/jjrt.2013_jsrt_69.9.952
Source DB: PubMed Journal: Nihon Hoshasen Gijutsu Gakkai Zasshi ISSN: 0369-4305