| Literature DB >> 24063963 |
Eric D Ryan1, Joseph G Rosenberg, Michael J Scharville, Eric J Sobolewski, Brennan J Thompson, Gilbert E King.
Abstract
The purpose of the present study was to examine test-retest reliability and minimal detectable change (MDC) values for Achilles tendon (AT) length determined using panoramic ultrasound (US) imaging. Seventeen men (age = 21.0 ± 2.3 y) visited the laboratory on two separate days, where AT length was examined along the mid-longitudinal axis of the right lower leg with a portable B-mode panoramic US device. These measures were found to have acceptable reliability with intra-class correlation coefficient (ICC) and standard error of measurement (SEM) values of 0.954 and 4.43 mm (SEM% of the mean = 2.37%), respectively. In addition, the MDC for the panoramic US assessment of AT length was 12.27 mm (MDC% of the mean = 6.57%). These findings suggest that panoramic US imaging is a reliable technique for detection of clinically relevant changes in AT length and may therefore be a practical and time-efficient clinical tool for future studies examining AT length in vivo.Entities:
Keywords: Extended field of view; Musculoskeletal; Sonography; Standard error of measurement
Mesh:
Year: 2013 PMID: 24063963 DOI: 10.1016/j.ultrasmedbio.2013.06.013
Source DB: PubMed Journal: Ultrasound Med Biol ISSN: 0301-5629 Impact factor: 2.998