MATERIAL AND METHODS: In order to investigate the prototype usability in experimental conditions, a biomechanical study was performed in which 42 polyacetal tubes set in 14 experimental groups and subgroups represented the fractured tibia that were fixed by a standard dynamic axial external fixator and a novel fixator. Displacements under static and dynamic loads were measured, with static ones corresponding to three directions of fragment movement and dynamic simulating the human gait. Analysis was performed in SPSS v13, with significance set at P<0.05.
RESULTS: The novel fixator showed biomechanical superiority in "fragments apart" study groups, while the standard dynamic axial external fixator outperformed the novel one in the situations of bending with "fragments in contact" study groups. There were no significant differences in dynamic load, despite better numerical result of the novel fixator.
CONCLUSION: The novel fixator is expectedly faster applicable and offers greater extent of external fixation flexibility. Further developments of this model thus seems justified in both construction improvement and on clinical application.
Copyright © 2013 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Biomechanics; Croatia; Cyclic load; Dynamic axial external fixator; Dynamic load; External fixator; Novel external fixator; Static load; Tibial fracture

Mesh:

Year:  2013        PMID: 24060020     DOI: 10.1016/S0020-1383(13)70199-2

Source DB:  PubMed          Journal:  Injury        ISSN: 0020-1383            Impact factor:   2.586


  1 in total

Review 1.  An engineering review of external fixators.

Authors:  P L N Fernando; Aravinda Abeygunawardane; Pci Wijesinghe; Parakrama Dharmaratne; Pujitha Silva
Journal:  Med Eng Phys       Date:  2021-11-04       Impact factor: 2.242

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.