Literature DB >> 24052795

Vascular steal explains early paradoxical blood oxygen level-dependent cerebrovascular response in brain regions with delayed arterial transit times.

Julien Poublanc1, Jay Show Han, Daniel Michael Mandell, John Conklin, Jeffrey Alan Stainsby, Joseph Arnold Fisher, David John Mikulis, Adrian Philip Crawley.   

Abstract

INTRODUCTION: Blood oxygen level-dependent (BOLD) magnetic resonance imaging (MRI) during manipulation of inhaled carbon dioxide (CO2) can be used to measure cerebrovascular reactivity (CVR) and map regions of exhausted cerebrovascular reserve. These regions exhibit a reduced or negative BOLD response to inhaled CO2. In this study, we sought to clarify the mechanism behind the negative BOLD response by investigating its time delay (TD). Dynamic susceptibility contrast (DSC) MRI with the injection of a contrast agent was used as the gold standard in order to provide measurement of the blood arrival time to which CVR TD could be compared. We hypothesize that if negative BOLD responses are the result of a steal phenomenon, they should be synchronized with positive BOLD responses from healthy brain tissue, even though the blood arrival time would be delayed.
METHODS: On a 3-tesla MRI system, BOLD CVR and DSC images were collected in a group of 19 patients with steno-occlusive cerebrovascular disease. For each patient, we generated a CVR magnitude map by regressing the BOLD signal with the end-tidal partial pressure of CO2 (PETCO2), and a CVR TD map by extracting the time of maximum cross-correlation between the BOLD signal and PETCO2. In addition, a blood arrival time map was generated by fitting the DSC signal with a gamma variate function. ROI masks corresponding to varying degrees of reactivity were constructed. Within these masks, the mean CVR magnitude, CVR TD and DSC blood arrival time were extracted and averaged over the 19 patients. CVR magnitude and CVR TD were then plotted against DSC blood arrival time.
RESULTS: The results show that CVR magnitude is highly correlated to DSC blood arrival time. As expected, the most compromised tissues with the longest blood arrival time have the lowest (most negative) CVR magnitude. However, CVR TD shows a noncontinuous relationship with DSC blood arrival time. CVR TD is well correlated to DSC blood arrival time (p < 0.0001) for tissue of positive reactivity, but fails to maintain this trend for tissue of negative reactivity. Regions with negative reactivity have similar CVR TD than healthy regions.
CONCLUSION: These results support the hypothesis that negative reactivity is the result of a steal phenomenon, lowering the BOLD signal as soon as healthier parts of the brain start to react and augment their blood flow. BOLD CVR MRI is capable of identifying this steal distribution, which has particular diagnostic significance as it represents an actual reduction in flow to already compromised tissue.

Entities:  

Keywords:  Blood oxygen level-dependent magnetic resonance imaging; Cerebrovascular reactivity; Dynamic contrast imaging; Steal phenomenon; Time delay

Year:  2013        PMID: 24052795      PMCID: PMC3776710          DOI: 10.1159/000348841

Source DB:  PubMed          Journal:  Cerebrovasc Dis Extra        ISSN: 1664-5456


Introduction

Blood oxygen level-dependent (BOLD) magnetic resonance imaging (MRI) can be used to measure cerebrovascular reactivity (CVR) to vasodilators such as carbon dioxide (CO2). In healthy brain tissue, increased end-tidal partial pressure of CO2 (PETCO2) causes arteriolar dilation, producing an increase in blood flow and an increased BOLD MRI signal. In patients with steno-occlusive disease, it is expected that compromised tissue will become chronically vasodilated and may exhibit an inability to further augment blood flow, which would lead to an absence of any BOLD response to a CO2 stimulus [1]. In fact, a significant decrease in BOLD signal is often seen in tissue that is presumed to be severely compromised [2]. This decrease in BOLD signal must correspond to an increase in the amount of deoxyhemoglobin within the voxel, which might be explained by either of two physiological models. We hypothesize that either a somewhat time-delayed volume increase occurs due to local vasodilation but the overall resistance including proximal stenoses does not allow for any significant increase in blood flow (model A) or other healthier tissue vasodilates before the compromised tissue under consideration and the immediate pressure drop causes an actual decrease in flow (model B). This second mechanism, named vascular steal phenomenon, has previously been described [3]. It occurs in brain regions with increased proximal resistance such as an arterial stenosis. In this case, the arterioles experience a lowered perfusion pressure and respond over time by a chronic compensatory vasodilation. This is effective in maintaining resting blood flow until maximal vasodilation is reached. At this point, if a vasodilator stimulus is delivered to the brain, the compromised tissue is unable to reduce its resistance, and blood flow is diverted to surrounding healthy tissues with preserved vasodilatory capacity. In this retrospective study, we attempt to distinguish those two models (A and B) on the basis of time delay (TD) measurements. We compare BOLD CVR scans with dynamic susceptibility contrast (DSC) imaging in a group of patients with steno-occlusive disease. The DSC technique uses an intravenous injection of a paramagnetic contrast agent (gadolinium) in order to induce a large MR signal drop via an intravascular T2* decrease. The relative arrival time of the gadolinium contrast in the capillary beds can be measured fairly accurately and provides a surrogate estimate for the relative arrival times of the arterial CO2 bolus that is created by modulation of the inhaled gas partial pressure of CO2 (PCO2). The true CVR response must occur after the arrival of the CO2 bolus. However, any change in pressure across the proximal resistance will occur as soon as the arterioles in the healthier tissue begin to dilate. Therefore, if hypothesis B is true, a negative CVR response due to vascular steal should occur in synchrony with the positive response of the healthier tissue causing the steal.

Methods

Study Population

The study protocol was approved by the University Health Network Research Ethics Board. A total of 19 patients (table 1) with steno-occlusive cerebrovascular disease, referred from the neurology and neurosurgery services at the Toronto Western Hospital, were recruited into this study. All patients provided written informed consent.
Table 1

Patient characteristics

Patient No./ Pathology
sex/age
1/M/15moyamoya, bilateral ICA stenosis
2/M/15moyamoya, severe right and mild left MCA stenosis
3/F/17moyamoya, left ICA stenosis
4/F/18cocaine-induced vasculitis
5/M26moyamoya, severe left MCA stenosis
6/M/36moyamoya, severe right MCA stenosis
7/M/45moyamoya, right ICA, right MCA and left MCA stenosis
8/M/46moyamoya, severe right MCA stenosis
9/F/48moyamoya, severe right MCA stenosis
10/M/56atherosclerosis and occlusion of the right ICA
11/F/60bilaterally vertebral artery occlusion, right ICA stenosis
12/M/60moyamoya, left ICA stenosis
13/M/69left ICA stenosis
14/M/71stroke, severe right ICA stenosis
15/F/74possible vasculopathy
16/M/74stroke, left MCA occlusion
17/M/75right ICA stenosis
18/F/83infarct in amygdala and thalamus of the right side
19/M/60severe right ICA stenosis

ICA = Internal carotid artery

MCA = middle cerebral artery.

Imaging Protocol

All patients were imaged on a 3-tesla MRI system (GE Medical Systems, Milwaukee, Wisc., USA), using an 8-channel head coil.

BOLD CVR

CVR images were acquired using a BOLD MR pulse sequence with the following parameters: TR = 2,000 ms, TE = 35 ms, flip angle = 85°, 28 slices, slice thickness = 4.5 mm, 64 × 64 matrix size, field of view = 200 mm, and either a spiral or echo planar imaging readout.

CVR Equipment and Protocol

Gas delivery to the patient was achieved using a custom-built automated gas blender and breathing circuit (RespirAct; Thornhill Research Inc., Toronto, Ont., Canada). This apparatus described by Slessarev et al. [4] enables the prospective control of the patient's PETCO2 and end-tidal pressure O2 (PETO2) independently of each other and of minute ventilation [5,6]. During BOLD MRI acquisition, patients were induced with alternating normocapnic (N) and hypercapnic (H) states during iso-oxia (PETO2 = 100 mm Hg). Normocapnic and hypercapnic states refer to PETCO2 = 40 mm Hg and PETCO2 = 50 mm Hg, respectively. The timing sequence was as follows: N/H/N/H/N = 45/45/90/150/180 s. Even though the above describes the most current material used in our institution, a more basic rebreathing system (where PETO2 is not controlled) was used in 13 patients. In these cases, normocapnic (40 mm Hg) and hypercapnic (50 mm Hg) states were alternating periodically for about 1 min [see [7] for details].

Anatomical Scan

For coregistration purposes, T1-weighted images were acquired using an inversion recovery-prepared spoiled gradient-recalled sequence with the following parameters: flip angle = 20°, TI/TE/TR = 300/5/11.5 ms, 60 slices of 2.2 mm thickness and a 256 × 256 matrix size.

DSC Imaging Scan

T2*-weighted dynamic images were acquired using gradient-echo, single-shot echo planar imaging (TR = 1,735 ms, TE = 31.5 ms, flip angle = 90°, 128 × 128 acquisition matrix, 20 × 20 cm field of view, 18 slices of 5 mm thickness and 2 mm gap) during a 15-ml intravenous injection of 7.5 mmol of gadodiamide (Omniscan) at a flow rate of 5 ml/s, immediately followed by a 30-ml saline flush.

Data Processing

End-Tidal Partial Pressure of CO2

PETCO2 was obtained by selecting the expiration peaks of the PCO2 curve and used as a reference waveform in the cross-correlation analysis.

MRI Data

All data were transferred to an independent workstation for further analysis. For each raw data set (CVR and DSC), slice-time correction, followed by volume re-registration for motion correction during scanning, was applied using the AFNI software [8]. Next, CVR and DSC volumes were realigned onto the anatomical image to correct for possible changes in the patient's head position between scans.

CVR Magnitude or Reactivity Map

To account for the unknown delay between the PETCO2 waveform and the actual arterial PCO2 in the brain, the PETCO2 waveform was first time-shifted to correspond to the maximum cross-correlation with the BOLD signal averaged over the brain. The BOLD time series at each voxel was then normalized and regressed against the PETCO2 waveform. The slope of this regression is the magnitude of the CVR signal and is expressed as a percentage change in BOLD signal per mm Hg (%/mm Hg) in PETCO2. The final CVR magnitude map (M; fig. 1a) is displayed with a graded color scale, and can be thresholded with the correlation coefficient resulting from the cross-correlation between the BOLD signal and PETCO2.
Fig. 1

CVR magnitude or M (a, d), TD (b, e) and BAT (c, f) maps for 2 patients. The first row (a-c) represents a patient with mainly positive reactivity, whereas the second row (d-f) represents a patient with an extended area of negative reactivity.

DSC Blood Arrival Time Map

AFNI was used to perform a nonlinear fitting procedure of the DSC time curve by a gamma variate of the form: S = (t − t0)a × exp(-(t − t0)/b), t0 being the start of the signal or blood arrival time (BAT). In this way, BAT was calculated for each voxel, and the resulting map (fig. 1b) was resampled to match the CVR grid.

CVR TD Map

An AFNI program called the Hilbert Delay [9,10] was used to calculate the TD maps (fig. 1c). This program estimates the time at which the cross-correlation between the PETCO2 and the BOLD signal time course is a maximum. The advantage of this method is its ability to generate TDs at a finer resolution than TR without having to entirely interpolate the signal and reference waveforms. It takes advantage of the fact that the cross-correlation is maximum as its Hilbert transform crosses zero. The zero crossing time point between two TRs is easily found with one linear interpolation. In this way, TD, along with a corresponding correlation value, was calculated on a voxel-by-voxel basis. Additionally, TD in voxels of negative reactivity, determined from the initial M map (M < 0), was calculated using an inverted PETCO2 waveform. The TD map obtained is relative to the PETCO2. Note that on a per patient basis, the position of the PETCO2 is not synchronized with the healthy brain but with the average brain signal. We later adjusted each patient's map to a common reference in order to group all 19 patients together. Also, it is important to realize that TD as defined above is very different than BAT. Due to the nature of the cross-correlation, it is a more general time combining a BAT as well as a response time. Response time could be defined as the time for the signal to go from baseline to maximum reactivity. TD differences between healthy and diseased tissue can be as long as half a minute in some patients, whereas corresponding BAT differences are on the order of a few seconds.

Data Analysis

For each patient, whole-brain CVR data were separated into multiple intervals or bins of M spanning 0.05%/mm Hg each and ranged from −0.20 to 0.60%/mm Hg. This method was used to construct 16 corresponding ROI masks. Additionally, as the program Hilbert Delay can potentially yield extremely long delays (physiologically impossible) for noisy time series, those masks were further restricted by excluding any voxel of TD greater than 45 s. For each patient, the number of voxels contained in each of the 16 ROI masks was used to construct the cumulative histogram (fig. 2). To reduce the influence of outliers caused by large veins (dural sinuses) or movement artifacts, we removed the highest 5% and the lowest 5% of all M values. As shown in figure 2, this defines the patient's range of reactivity, with bmin and bmax being the minimum and maximum bin included.
Fig. 2

Cumulative histogram of reactivity (M) for 1 patient. It is used to define the patient range of reactivity. The highest 5% and the lowest 5% of all M values are excluded from the analysis (white bins). In dark and light grey are all the included bins for this patient. In dark grey are the minimum and maximum reactivity bins (bmin and bmax, respectively).

In each patient, average M, TD and BAT were calculated for each of the corresponding ROI masks included in the reactivity range. Since TD and BAT are not absolute parameters, a common reference between all patients must be defined. We assume that healthy tissue with maximum reactivity should have the earliest TD and BAT. Consequently, for each patient separately, we used the ROI mask of bmax as the zero reference and shifted all TD and BAT accordingly. It was then possible to average M, TD and BAT across patients for each separate bin. Note that the number of patients contributing to different ROI masks (constructed from reactivity intervals) can vary as the individual range of reactivity varies from patient to patient. Final averaged M and TD were plotted against averaged BAT (fig. 2).

Results

Figure 1 shows the CVR M, TD and BAT for 2 patients. The first patient shows a good correspondence between positive diminished reactivity (fig. 1a) and prolonged TD (fig. 1b) and BAT (fig. 1c). In the second patient, the region of negative reactivity (fig. 1d) has a prolonged BAT (fig. 1f) but a short TD (fig. 1e). Figure 3a shows CVR M plotted against DSC BAT. Black and grey data points represent the positive and negative reactivity data, respectively. Each of the data points corresponds to one of the bins. Next to each data point is the number of patients contributing to the average M and BAT (grey text). This graph clearly shows that M and BAT are related.
Fig. 3

CVR magnitude (M) plotted against DSC BAT (a) as well as CVR TD plotted against DSC BAT (b) for all patients. Positive and negative reactivity is represented, respectively, in black and grey. a The number of patients represented by each data point is shown in grey.

Figure 3b shows CVR TD plotted against DSC BAT. Regression was used to analyze these data, with the constraint of BAT and TD being simultaneously null in healthy tissue. If all black and grey data points are taken into account, TD is not correlated to BAT (n = 16, R2 = 0.09, p = 0.27). However, if only data from regions of positive reactivity (M) are considered (black data points), TD is well correlated to BAT (n = 12, R2 = 0.88, p < 0.0001). As expected (see Methods section) TD is much longer than BAT. Note that there is a 10-fold difference between the two parameters.

Discussion

The conventional parameter extracted from a BOLD CVR experiment is the magnitude of the BOLD response expressed as a percent BOLD signal increase per mm Hg PETCO2. Figure 3a shows that this parameter, M, is well correlated to the BAT parameter estimated by the DSC technique. This implies that the two techniques detect vascular impairment in similar brain regions. However, we observed that healthy tissue with no BAT delay (BAT ≈ 0) has a surprisingly wide range of reactivity (M ≈ 0.4-0.6%/mm Hg) illustrated by the steep descending portion of the curve in figure 3a. This may reflect partial volume effects in voxels containing large draining veins. As those voxels have a greater blood volume than voxels containing only tissue, deoxyhemoglobin content and subsequently BOLD signal will be much larger. The steepness of the curve around zero BAT would decrease if large draining veins were masked out. This effect was reduced by defining the patient reactivity range illustrated in figure 2. In contrast to the continuous relationship between M and BAT seen in figure 3a, TD plotted against BAT (fig. 3b) shows that voxels with negative reactivity (grey data points) behave very differently from voxels with positive reactivity (black data points). From figure 3a, we know that the voxels with negative reactivity correspond to the most impaired vasculature because the reference technique (DSC) shows that they experience the longest BAT. However, these voxels (grey data points) are not detected as impaired vasculature by CVR TD. Indeed, TD increases linearly with BAT for positive M, but this trend fails spectacularly when M becomes negative. The grey data points have the lowest TD, which conflicts with any expectation that increasing TD may reflect increased severity of disease, and this would be consistent with hypothesis A. We interpret this result as evidence for hypothesis B, i.e. that negative reactivity is the result of a steal phenomenon, which lowers the BOLD signal in severely diseased tissue as soon as healthier parts of the brain start to react and augment their blood flow. Our observation that negative reactivity is secondary to a steal phenomenon is supported by previous work in subjects with high-grade vascular stenosis [11]. In addition, arterial spin labeling, which provides a more direct measure of blood flow, indicates a reduced blood flow in tissue that shows a reduced BOLD CVR magnitude [12]. The vasculature in regions with negative ‘reactivity’ is not reacting to a local increase in PETCO2. Rather, the reduction in blood flow is driven passively by the dilation of vessels in the surrounding healthier tissue. The DSC data demonstrate that the gadolinium bolus and equivalently the CO2 bolus usually arrive at the more compromised tissue slightly later than at the surrounding healthier tissue. As the healthier tissue begins to vasodilate, a greater pressure drop across the stenosed arteries will reduce the perfusion pressure simultaneously at all downstream arterioles. We expect the BOLD signal from severely compromised tissue that is susceptible to a steal effect to start to change in exact antiphase to the signal from the normally reacting healthier tissue. However, as observed in figure 3b, three of the grey data points have a shorter TD than the reference TD (TD = 0 for Mmax) in healthy tissue. In order to understand this paradoxical result, it is important to realize that the negative BOLD signal is not an exact reflection (about the time axis) of the positive reactivity found in healthy vasculature (fig. 4). Visual examination of a negative BOLD time series during the hypercapnic stage shows that in some cases a slow increase in signal can follow the initial drop (fig. 4, grey line). We conclude that the negative BOLD signal is in general a combination of a steal phenomenon with a small delayed local (positive) reactivity. These two effects introduce an asymmetry in the signal shape and shift the TD calculation toward the initial drop, where the steal phenomenon dominates. This artifact explains how a few (grey) data points can have negative TD.
Fig. 4

The raw BOLD signal for 2 voxels from the same patient is shown. One voxel demonstrates positive reactivity to CO2 (black line) while the other demonstrates negative reactivity (grey line). The PETCO2 (dashed line) varies between approximately 40 and 50 mm Hg. The grey signal shows an initial drop as soon as PCO2 is elevated. However, this is followed by a signal increase during the high PCO2 plateau, which we interpret as a combination of some residual component of true reactivity (vasodilation of vascular beds within this voxel) coupled with steal phenomenon as described in the text.

In this study, the TD parameter was only used to confirm the existence of cerebral steal physiology but could provide useful diagnostic information regarding the nature of the hemodynamic impairment. In particular, TD could be used to locate collateral circulation, which is very important for flow compensation in patients with intracranial stenosis [13]. Also, it should be noted that a previous study [14] using transcranial Doppler found that the delay between the decrease in PETCO2 and the subsequent decrease in cerebral blood flow velocity correlated with cerebral hemodynamic compromise. However, extracting time information from a low signal-to-noise time series such as BOLD CVR is challenging especially in impaired vasculature with reactivity approaching zero. Furthermore, we now know that the timing of a negative BOLD signal does not reflect local response time to the vasodilatory stimulus. Moreover, this parameter can normally only be interpreted relative to surrounding areas (in this work, TD was expressed relative to the TD for healthy tissue). In contrast, the CVR magnitude does not have these disadvantages. First, M is stable even in low reactivity regions, since it is simply a measure of the amplitude of the signal. Second, M is an absolute measure at each voxel, and is scaled per mm Hg, making the maps comparable between patients.

Conclusion

Despite their fundamental difference, BOLD CVR magnitude maps and DSC BAT maps detected vascular impairment in similar locations. Comparison of CVR TDs with DSC BAT confirmed the existence of a steal phenomenon which explains the apparently paradoxical nature of a negative BOLD response to hypercapnia in severely compromised tissue. A negative BOLD response might be considered particularly diagnostically significant, since it does represent an actual reduction in flow to already compromised tissue, at least immediately following a global hypercapnia challenge. The point at which steal occurs depends upon the resistance of the stenosed proximal vessels and the resistance of the tissue that demonstrates the flow reduction. In a sense, the global stimulus mimics an ischemic event for the tissue at risk. BOLD CVR MRI is capable of identifying this steal distribution, which can be viewed as preischemic risk imaging. The clinical relevance of identifying this tissue at risk is the severalfold higher incidence of ischemic injury in patients with this physiology [15,16,17,18].

Disclosure Statement

Two of the study authors (J.A.F. and D.J.M.) contributed to the development of RespirAct™. These authors stand to gain financially if the device is successfully commercialized by Thornhill Research Inc., a University of Toronto/University Health Network-related company.
  18 in total

1.  Analysis and use of FMRI response delays.

Authors:  Z S Saad; K M Ropella; R W Cox; E A DeYoe
Journal:  Hum Brain Mapp       Date:  2001-06       Impact factor: 5.038

2.  MRI mapping of cerebrovascular reactivity using square wave changes in end-tidal PCO2.

Authors:  A Vesely; H Sasano; G Volgyesi; R Somogyi; J Tesler; L Fedorko; J Grynspan; A Crawley; J A Fisher; D Mikulis
Journal:  Magn Reson Med       Date:  2001-06       Impact factor: 4.668

3.  Precise control of end-tidal carbon dioxide levels using sequential rebreathing circuits.

Authors:  R B Somogyi; A E Vesely; D Preiss; E Prisman; G Volgyesi; T Azami; S Iscoe; J A Fisher; H Sasano
Journal:  Anaesth Intensive Care       Date:  2005-12       Impact factor: 1.669

4.  Motor-related intracortical steal phenomenon detected by multichannel functional near-infrared spectroscopy imaging.

Authors:  Takenori Akiyama; Takayuki Ohira; Toshinori Kato; Yasuo Toda; Maaya Orii; Kenji Hiraga; Atsushi Fukunaga; Masahito Kobayashi; Satoshi Onozuka; Takeshi Kawase
Journal:  Cerebrovasc Dis       Date:  2005-08-30       Impact factor: 2.762

5.  Correlation of cerebrovascular reserve as measured by acetazolamide-challenged SPECT with angiographic flow patterns and intra- or extracranial arterial stenosis.

Authors:  H T Ozgur; T Kent Walsh; A Masaryk; J F Seeger; W Williams; E Krupinski; M Melgar; E Labadie
Journal:  AJNR Am J Neuroradiol       Date:  2001-05       Impact factor: 3.825

6.  Mapping cerebrovascular reactivity using blood oxygen level-dependent MRI in Patients with arterial steno-occlusive disease: comparison with arterial spin labeling MRI.

Authors:  Daniel M Mandell; Jay S Han; Julien Poublanc; Adrian P Crawley; Jeff A Stainsby; Joseph A Fisher; David J Mikulis
Journal:  Stroke       Date:  2008-05-01       Impact factor: 7.914

7.  Time delay as a parameter for cerebrovascular reactivity in patients with severe carotid stenosis.

Authors:  A Hetzel; B Guschlbauer; M Reinhard
Journal:  Cerebrovasc Dis       Date:  2003       Impact factor: 2.762

8.  Increased stroke risk predicted by compromised cerebral blood flow reactivity.

Authors:  H Yonas; H A Smith; S R Durham; S L Pentheny; D W Johnson
Journal:  J Neurosurg       Date:  1993-10       Impact factor: 5.115

9.  Prospective targeting and control of end-tidal CO2 and O2 concentrations.

Authors:  Marat Slessarev; Jay Han; Alexandra Mardimae; Eitan Prisman; David Preiss; George Volgyesi; Cliff Ansel; James Duffin; Joseph A Fisher
Journal:  J Physiol       Date:  2007-04-19       Impact factor: 5.182

10.  Comparison of the effects of independently-controlled end-tidal PCO(2) and PO(2) on blood oxygen level-dependent (BOLD) MRI.

Authors:  Eitan Prisman; Marat Slessarev; Jay Han; Julien Poublanc; Alexandra Mardimae; Adrian Crawley; Joseph Fisher; David Mikulis
Journal:  J Magn Reson Imaging       Date:  2008-01       Impact factor: 4.813

View more
  21 in total

1.  Assessing cerebrovascular reactivity by the pattern of response to progressive hypercapnia.

Authors:  Joseph A Fisher; Olivia Sobczyk; Adrian Crawley; Julien Poublanc; Paul Dufort; Lashmi Venkatraghavan; Kevin Sam; David Mikulis; James Duffin
Journal:  Hum Brain Mapp       Date:  2017-04-03       Impact factor: 5.038

2.  Snapshot MR technique to measure OEF using rapid frequency mapping.

Authors:  Rajiv G Menon; Edward G Walsh; Donald B Twieg; Charles G Cantrell; Parmede Vakil; Sumeeth V Jonathan; Hunt H Batjer; Timothy J Carroll
Journal:  J Cereb Blood Flow Metab       Date:  2014-04-23       Impact factor: 6.200

3.  Transient Hypoxia Model Revealed Cerebrovascular Impairment in Anemia Using BOLD MRI and Near-Infrared Spectroscopy.

Authors:  Julie Coloigner; Chau Vu; Matthew Borzage; Adam Bush; Soyoung Choi; Xin Miao; Yaqiong Chai; Cristina Galarza; Natasha Lepore; Benita Tamrazi; Thomas D Coates; John C Wood
Journal:  J Magn Reson Imaging       Date:  2020-07-09       Impact factor: 4.813

4.  The role of vascular resistance in BOLD responses to progressive hypercapnia.

Authors:  James Duffin; Olivia Sobczyk; Adrian Crawley; Julien Poublanc; Lashmi Venkatraghavan; Kevin Sam; Alan Mutch; David Mikulis; Joseph Fisher
Journal:  Hum Brain Mapp       Date:  2017-08-07       Impact factor: 5.038

5.  The effects of perturbed cerebral blood flow and cerebrovascular reactivity on structural MRI and behavioral readouts in mild traumatic brain injury.

Authors:  Justin A Long; Lora T Watts; Wei Li; Qiang Shen; Eric R Muir; Shiliang Huang; Robert C Boggs; Abhinav Suri; Timothy Q Duong
Journal:  J Cereb Blood Flow Metab       Date:  2015-06-24       Impact factor: 6.200

6.  Measuring cerebrovascular reactivity: the dynamic response to a step hypercapnic stimulus.

Authors:  Julien Poublanc; Adrian P Crawley; Olivia Sobczyk; Gaspard Montandon; Kevin Sam; Daniel M Mandell; Paul Dufort; Lashmikumar Venkatraghavan; James Duffin; David J Mikulis; Joseph A Fisher
Journal:  J Cereb Blood Flow Metab       Date:  2015-07-01       Impact factor: 6.200

7.  Systematic protocol for assessment of the validity of BOLD MRI in a rabbit model of inflammatory arthritis at 1.5 tesla.

Authors:  Michael W Chan; George Nathanael; Antonella Kis; Afsaneh Amirabadi; Anguo Zhong; Tammy Rayner; Ruth Weiss; Garry Detzler; Roland Jong; Harpal Gahunia; Rahim Moineddin; Adrian Crawley; Andrea S Doria
Journal:  Pediatr Radiol       Date:  2013-12-24

8.  Quantitative cerebrovascular pathology in a community-based cohort of older adults.

Authors:  Swati Rane; Natalie Koh; Peter Boord; Tara Madhyastha; Mary K Askren; Suman Jayadev; Brenna Cholerton; Eric Larson; Thomas J Grabowski
Journal:  Neurobiol Aging       Date:  2018-01-31       Impact factor: 4.673

9.  On the optimization of imaging protocol for the mapping of cerebrovascular reactivity.

Authors:  Harshan Ravi; Binu P Thomas; Shin-Lei Peng; Hanli Liu; Hanzhang Lu
Journal:  J Magn Reson Imaging       Date:  2015-08-13       Impact factor: 4.813

10.  BOLD delay times using group delay in sickle cell disease.

Authors:  Julie Coloigner; Chau Vu; Adam Bush; Matt Borzage; Vidya Rajagopalan; Natasha Lepore; John Wood
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2016-03-21
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.