Literature DB >> 24047533

Scoring accuracy of automated sleep staging from a bipolar electroocular recording compared to manual scoring by multiple raters.

Carl Stepnowsky1, Daniel Levendowski, Djordje Popovic, Indu Ayappa, David M Rapoport.   

Abstract

OBJECTIVES: Electroencephalography (EEG) assessment in research and clinical studies is limited by the patient burden of multiple electrodes and the time needed to manually score records. The objective of our study was to investigate the accuracy of an automated sleep-staging algorithm which is based on a single bipolar EEG signal.
METHODS: Three raters each manually scored the polysomnographic (PSG) records from 44 patients referred for sleep evaluation. Twenty-one PSG records were scored by Rechtschaffen and Kales (R&K) criteria (group 1) and 23 PSGs were scored by American Academy of Sleep Medicine (AASM) 2007 criteria (group 2). Majority agreement was present in 98.4% of epochs and was used for comparison to automated scoring from a single EEG lead derived from the left and right electrooculogram.
RESULTS: The κ coefficients for interrater manual scoring ranged from 0.46 to 0.89. The κ coefficient for the auto algorithm vs manual scoring by rater ranged from 0.42 to 0.63 and was 0.61 (group 1, κ=0.61 and group 2, κ=0.62) for majority agreement for all studies. The mean positive percent agreement across subjects and stages was 72.6%, approximately 80% for stages wake (78.3%), stage 2 sleep (N2) (80.9%), and stage 3 sleep (N3) (78.1%); the percentage slightly decreased to 73.2% for rapid eye movement (REM) sleep and dropped to 31.9% for stage 1 sleep (N1). Differences in agreement were observed based on raters, obstructive sleep apnea (OSA) severity, medications, and signal quality.
CONCLUSIONS: Our study demonstrated that automated scoring of sleep obtained from a single-channel of forehead EEG results in agreement to majority manual scoring are similar to results obtained from studies of manual interrater agreement. The benefit in assessing auto-staging accuracy with consensus agreement across multiple raters is most apparent in patients with OSA; additionally, assessing auto-staging accuracy limited disagreements in patients on medications and in those with compromised signal quality. Published by Elsevier B.V.

Entities:  

Keywords:  Automatic sleep scoring; Electroencephalography; Electrooculography; Polysomnography; Sleep stages; Validation studies

Mesh:

Year:  2013        PMID: 24047533     DOI: 10.1016/j.sleep.2013.04.022

Source DB:  PubMed          Journal:  Sleep Med        ISSN: 1389-9457            Impact factor:   3.492


  34 in total

1.  Electro-oculography-based detection of sleep-wake in sleep apnea patients.

Authors:  Jussi Virkkala; Jussi Toppila; Paula Maasilta; Adel Bachour
Journal:  Sleep Breath       Date:  2014-10-01       Impact factor: 2.816

2.  A State Space and Density Estimation Framework for Sleep Staging in Obstructive Sleep Apnea.

Authors:  Dae Y Kang; Pamela N DeYoung; Atul Malhotra; Robert L Owens; Todd P Coleman
Journal:  IEEE Trans Biomed Eng       Date:  2017-05-08       Impact factor: 4.538

3.  Performance of a New Portable Wireless Sleep Monitor.

Authors:  Magdy Younes; Marc Soiferman; Wayne Thompson; Eleni Giannouli
Journal:  J Clin Sleep Med       Date:  2017-02-15       Impact factor: 4.062

4.  Validation of a Wireless, Self-Application, Ambulatory Electroencephalographic Sleep Monitoring Device in Healthy Volunteers.

Authors:  Patrick H Finan; Jessica M Richards; Charlene E Gamaldo; Dingfen Han; Jeannie Marie Leoutsakos; Rachel Salas; Michael R Irwin; Michael T Smith
Journal:  J Clin Sleep Med       Date:  2016-11-15       Impact factor: 4.062

5.  Accuracy of Automatic Polysomnography Scoring Using Frontal Electrodes.

Authors:  Magdy Younes; Mark Younes; Eleni Giannouli
Journal:  J Clin Sleep Med       Date:  2016-05-15       Impact factor: 4.062

6.  Odds ratio product of sleep EEG as a continuous measure of sleep state.

Authors:  Magdy Younes; Michele Ostrowski; Marc Soiferman; Henry Younes; Mark Younes; Jill Raneri; Patrick Hanly
Journal:  Sleep       Date:  2015-04-01       Impact factor: 5.849

7.  Comparison of two home sleep testing devices with different strategies for diagnosis of OSA.

Authors:  Tyler Gumb; Akosua Twumasi; Shahnaz Alimokhtari; Alan Perez; Kathleen Black; David M Rapoport; Jag Sunderram; Indu Ayappa
Journal:  Sleep Breath       Date:  2017-08-19       Impact factor: 2.816

8.  Addition of frontal EEG to adult home sleep apnea testing: does a more accurate determination of sleep time make a difference?

Authors:  Matthew P Light; Thalia N Casimire; Catherine Chua; Viachaslau Koushyk; Omar E Burschtin; Indu Ayappa; David M Rapoport
Journal:  Sleep Breath       Date:  2018-10-11       Impact factor: 2.816

9.  Automatic Sleep Staging in Patients With Obstructive Sleep Apnea Using Single-Channel Frontal EEG.

Authors:  Pei-Lin Lee; Yi-Hao Huang; Po-Chen Lin; Yu-An Chiao; Jen-Wen Hou; Hsiang-Wen Liu; Ya-Ling Huang; Yu-Ting Liu; Tzi-Dar Chiueh
Journal:  J Clin Sleep Med       Date:  2019-10-15       Impact factor: 4.062

10.  Comparison of a single-channel EEG sleep study to polysomnography.

Authors:  Brendan P Lucey; Jennifer S Mcleland; Cristina D Toedebusch; Jill Boyd; John C Morris; Eric C Landsness; Kelvin Yamada; David M Holtzman
Journal:  J Sleep Res       Date:  2016-06-02       Impact factor: 3.981

View more