Literature DB >> 24044977

Myocardial deformation analysis in contrast echocardiography: first results using two-dimensional cardiac performance analysis.

Alda Huqi1, Allen He, Berthold Klas, Ian Paterson, Richard Thompson, Marleen Irwin, Justin Ezekowitz, Jonathan B Choy, Harald Becher.   

Abstract

BACKGROUND: Contrast echocardiography (CE) provides closer agreement with magnetic resonance imaging (MRI) for left ventricular (LV) volumes and ejection fraction (EF) than noncontrast echocardiography. However, the feasibility and role of myocardial deformation analysis on contrast echocardiographic images have not been well established. The aim of this study was to assess the feasibility of deformation analysis on CE using a new software tool that provides simultaneous measurements for LV volumes and EF.
METHODS: Data from 52 patients who were recruited for the Alberta Heart Failure Etiology and Analysis Research Team Study (34 men; mean age, 64 ± 9 years) and underwent CE and MRI were considered. Contrast bolus injections were administered for optimal endocardial definition. Offline LV volume analysis was performed by standard manual tracing. A single frame was traced manually for two-dimensional (2D) cardiac performance analysis (CPA), which automatically calculated LV volumes, EF, and global longitudinal strain (GLS). Volumes obtained with 2D CPA were compared with those measured with standard CE and MRI. GLS from noncontrast echocardiographic recordings was also calculated with 2D CPA and compared with CE-derived and MRI-derived GLS.
RESULTS: Tracing of contrast echocardiographic images with 2D CPA was possible in 49 out of 52 patients, and measurements correlated well with standard CE and MRI (EF: r = 0.93, P < .001, and r = 0.85, P < .001, respectively). Mean GLS from noncontrast echocardiographic and contrast echocardiographic recordings was -13.4 ± 5.8 and -15.3 ± 4.64, respectively (P = .056), and the latter correlated well with MRI-derived GLS (r = 0.78 vs 0.81, respectively).
CONCLUSIONS: Simultaneous volumetric and deformation analysis on contrast echocardiographic recordings is feasible and reproducible. While volumes and EF obtained with the new software compare well with those obtained from standard CE and MRI, GLS from CE shows a good correlation with strain measured with MRI.
Copyright © 2013 American Society of Echocardiography. Published by Mosby, Inc. All rights reserved.

Entities:  

Keywords:  2D; CE; CPA; Cardiac magnetic resonance; Cardiac performance analysis; Contrast echocardiography; EF; ESV; Ejection fraction; End-systolic volume; GLS; Global longitudinal strain; HF; Heart failure; LV; Left ventricular; MRI; Magnetic resonance imaging; Speckle-tracking analysis; Two-dimensional; Volumetric analysis

Mesh:

Substances:

Year:  2013        PMID: 24044977     DOI: 10.1016/j.echo.2013.08.010

Source DB:  PubMed          Journal:  J Am Soc Echocardiogr        ISSN: 0894-7317            Impact factor:   5.251


  2 in total

1.  Shenzhu Guanxin Recipe Granules () for Improving Exercise Tolerance in Patients with Stable Angina (SERIES Trial): A Protocol of Multicenter, Randomized, Double-Blind, Placebo Parallel Controlled Clinical Trial.

Authors:  Shuai Mao; Dan-Ping Xu; Xiao-Jing Dang; Winny Li; Huan-Lin Wu
Journal:  Chin J Integr Med       Date:  2018-10-17       Impact factor: 1.978

2.  Echocardiographic assessment of myocardial function and mechanics during veno-venous extracorporeal membrane oxygenation.

Authors:  David G Platts; Kenji Shiino; Jonathan Chan; Darryl J Burstow; Gregory M Scalia; John F Fraser
Journal:  Echo Res Pract       Date:  2019-06-01
  2 in total