A comparative study of the peripheral doses from a linear accelerator with a multileaf collimator system.

Hediye Acun1, Ali Zubaroglu, Gönül Kemikler, Ahmet Bozkurt.   

Abstract

This study presents a comparison of peripheral doses (PDs) measured using an ionisation chamber with treatment planning system (TPS) data and a Monte Carlo (MC) simulation of a 6-MV photon beam. The ion chamber measurements and MC simulation produced similar results for all out-of-field distances and field sizes considered in this study. For the 0° and 90° collimation angles, the average local per cent dose differences between the MC and TPS calculations were 2.7 % (range: -2.4, +22.6) and -1.7 % (range: -12.2, +10.8), respectively. The corresponding differences between the MC calculations and the ion chamber measurements were 2.2 % (range: -2.4, 24.7) and -1.8 % (range: -17, 15.2) for all field sizes and depths, respectively. Whereas the PDs increased with field sizes, the variations with depth were negligible at large distances. The TPS calculations usually yielded higher PDs than ion chamber measurements at distances close to the field edge. In contrast, at the farther distances, the TPS results indicated lower doses than both the ion chamber and the MC data. TPS data are not sufficient for use in calculating the out-of-field doses. These results can be used to estimate non-target organ doses to patients.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24043874     DOI: 10.1093/rpd/nct225

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  1 in total

1.  A Phantom Study on Fetal Dose Reducing Factors in Pregnant Patients with Breast Cancer during Radiotherapy Treatment.

Authors:  Akın Öğretici; Aydın Çakır; Uğur Akbaş; Canan Köksal; Ümmühan Kalafat; Makbule Tambaş; Hatice Bilge
Journal:  J Med Phys       Date:  2017 Jul-Sep
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.