PURPOSE: The objective of this study is to examine the effect of different firmware versions on ActiGraph™ counts from the laboratory, field, and mechanical shaker testing.
METHODS: Counts from 5 GT3X and 7 GT1M firmware versions were compared in this study. Monitors uploaded with these firmware versions were worn on the hip by 15 participants (age = 24.9 ± 5.0 yr, BMI= 23.9 ± 2.4 kg · m(-2)) who performed laboratory-based treadmill (walking: 1.5, 3.0 and 4.5 mph; running: 6 mph) and simulated free-living activities (sitting, self-paced walking, filing papers, dusting, vacuuming, and cleaning the room). Testing was also conducted during 1 d of free living and using orbital mechanical shaker testing at 0.7, 1.3, 2.0, and 3.0 Hz. Intermonitor comparisons for vertical, anteroposterior, mediolateral, and triaxial vector magnitude counts were conducted using one-way ANOVA and post hoc pairwise comparisons (P < 0.05).
RESULTS: Vertical counts during treadmill walking at 1.5 mph from the GT1M monitor with firmware version 1.1.0. were significantly greater (P < 0.05; 75%) than output from the monitor with firmware version 1.3.0. Shaker testing revealed statistically significant differences in vertical and lateral counts. Although there were no significant differences among activity counts in the free-living comparisons, firmware version 1.1.0. produced the highest vertical counts during this protocol.
CONCLUSION: Greater sensitivity of firmware version 1.1.0. to low-frequency sedentary activities resulted in greater counts than other firmware versions. It is recommended that before releasing new firmware, ActiGraph™ perform both human and mechanical shaker testing to verify comparability in outputs between new and previous firmware versions.

Entities:  

Mesh:

Year:  2014        PMID: 24042309     DOI: 10.1249/MSS.0000000000000145

Source DB:  PubMed          Journal:  Med Sci Sports Exerc        ISSN: 0195-9131            Impact factor:   5.411


  6 in total

Review 1.  The Influence of Head Impact Threshold for Reporting Data in Contact and Collision Sports: Systematic Review and Original Data Analysis.

Authors:  D King; P Hume; C Gissane; M Brughelli; T Clark
Journal:  Sports Med       Date:  2016-02       Impact factor: 11.136

2.  An Open-Source Monitor-Independent Movement Summary for Accelerometer Data Processing.

Authors:  Dinesh John; Qu Tang; Fahd Albinali; Stephen Intille
Journal:  J Meas Phys Behav       Date:  2019-12

3.  Recommendations for determining the validity of consumer wearable heart rate devices: expert statement and checklist of the INTERLIVE Network.

Authors:  Jan M Mühlen; Julie Stang; Esben Lykke Skovgaard; Pedro B Judice; Pablo Molina-Garcia; William Johnston; Luís B Sardinha; Francisco B Ortega; Brian Caulfield; Wilhelm Bloch; Sulin Cheng; Ulf Ekelund; Jan Christian Brønd; Anders Grøntved; Moritz Schumann
Journal:  Br J Sports Med       Date:  2021-01-04       Impact factor: 13.800

Review 4.  Systematic review of the validity and reliability of consumer-wearable activity trackers.

Authors:  Kelly R Evenson; Michelle M Goto; Robert D Furberg
Journal:  Int J Behav Nutr Phys Act       Date:  2015-12-18       Impact factor: 6.457

5.  The Validity of the Energy Expenditure Criteria Based on Open Source Code through two Inertial Sensors.

Authors:  Jaime Martín-Martín; Li Wang; Irene De-Torres; Adrian Escriche-Escuder; Manuel González-Sánchez; Antonio Muro-Culebras; Cristina Roldán-Jiménez; María Ruiz-Muñoz; Fermín Mayoral-Cleries; Attila Biró; Wen Tang; Borjanka Nikolova; Alfredo Salvatore; Antonio I Cuesta-Vargas
Journal:  Sensors (Basel)       Date:  2022-03-26       Impact factor: 3.576

6.  Youth Oriented Activity Trackers: Comprehensive Laboratory- and Field-Based Validation.

Authors:  John R Sirard; Brittany Masteller; Patty S Freedson; Albert Mendoza; Amanda Hickey
Journal:  J Med Internet Res       Date:  2017-07-19       Impact factor: 5.428

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.