High-dose-rate prostate brachytherapy based on registered transrectal ultrasound and in-room cone-beam CT images.

Aniek J G Even1, Tonnis T Nuver2, Hendrik Westendorp3, Carel J Hoekstra4, Cornelis H Slump5, André W Minken3.   

Abstract

PURPOSE: To present a high-dose-rate (HDR) brachytherapy procedure for prostate cancer using transrectal ultrasound (TRUS) to contour the regions of interest and registered in-room cone-beam CT (CBCT) images for needle reconstruction. To characterize the registration uncertainties between the two imaging modalities and explore the possibility of performing the procedure solely on TRUS. METHODS AND MATERIALS: Patients were treated with a TRUS/CBCT-based HDR brachytherapy procedure. For 100 patients, dosimetric results were analyzed. For 40 patients, registration uncertainties were examined by determining differences in fiducial marker positions on TRUS and registered CBCT. The accuracy of needle reconstruction on TRUS was investigated by determining the position differences of needle tips on TRUS and CBCT. The dosimetric impact of reregistration and needle reconstruction on TRUS only was studied for 8 patients.
RESULTS: The average prostate V100 was 97.8%, urethra D10 was 116.3%, and rectum D1 cc was 66.4% of the prescribed dose. For 85% of the patients, registration inaccuracies were within 3 mm. Large differences were found between needle tips on TRUS and CBCT, especially in cranial-caudal direction, with a maximum of 10.4 mm. Reregistration resulted in a maximum V100 reduction of 0.9%, whereas needle reconstruction on TRUS only gave a maximum reduction of 9.4%.
CONCLUSIONS: HDR prostate brachytherapy based on TRUS combined with CBCT is an accurate method. Registration uncertainties, and consequently dosimetric inaccuracies, are small compared with the uncertainties of performing the procedure solely based on static TRUS images. CBCT imaging is a requisite in our current procedure.
Copyright © 2014 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Cone-beam CT; HDR brachytherapy; Prostate; Registration; Transrectal ultrasound

Mesh:

Year:  2013        PMID: 24041955     DOI: 10.1016/j.brachy.2013.08.001

Source DB:  PubMed          Journal:  Brachytherapy        ISSN: 1538-4721            Impact factor:   2.362


  4 in total

1.  Prostate CT segmentation method based on nonrigid registration in ultrasound-guided CT-based HDR prostate brachytherapy.

Authors:  Xiaofeng Yang; Peter Rossi; Tomi Ogunleye; David M Marcus; Ashesh B Jani; Hui Mao; Walter J Curran; Tian Liu
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

2.  Evaluation of a mobile C-arm cone-beam CT in interstitial high-dose-rate prostate brachytherapy treatment planning.

Authors:  Mario Djukelic; David Waterhouse; Ryan Toh; Hendrick Tan; Pejman Rowshanfarzad; David Joseph; Martin A Ebert
Journal:  J Med Radiat Sci       Date:  2019-04-03

3.  Use of deformable image registration techniques to estimate dose to organs at risk following prostate external beam radiation therapy and high-dose-rate brachytherapy.

Authors:  Marie Vozzo; Joel Poder; Johnson Yuen; Joseph Bucci; Annette Haworth
Journal:  J Contemp Brachytherapy       Date:  2021-02-18

4.  Single-fraction high-dose-rate brachytherapy using real-time transrectal ultrasound based planning in combination with external beam radiotherapy for prostate cancer: dosimetrics and early clinical results.

Authors:  Olivier Lauche; Guila Delouya; Daniel Taussky; Cynthia Menard; Dominic Béliveau-Nadeau; Yannick Hervieux; Renée Larouche; Maroie Barkati
Journal:  J Contemp Brachytherapy       Date:  2016-04-14
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.