Abstract
The performance values of available techniques used in serodiagnosis of toxoplasmosis are satisfactory but they raise problems of equivocal and discordant results for very low IgG titers. Recently marketed, LDBio-Toxo II IgG Western blot (IB) showed an excellent correlation with the dye test. We estimated the proportion of equivocal and discordant results between the enzyme immunoassay Platelia Toxo IgG (EIA-IgG) and fluorescent antibody test (FAT) and assessed the usefulness of the IB as a confirmatory test. Out of 2,136 sera collected from pregnant women, 1,644 (77.0%) tested unequivocally positive and 407 (19.0%) were negative in both EIA-IgG and FAT. The remaining 85 (4%) sera showed equivocal or discordant results. Among them, 73 (85.9%) were positive and 12 (14.1%) were negative in IB. Forty-one (89.1%) equivocal sera in EIA-IgG and 46 (86.8%) equivocal sera in FAT were positive in IB. Reducing the cut-off values of both screening techniques improved significantly their sensitivity in detecting very low IgG titers at the expense of their specificity. In conclusion, equivocal results in routine-used techniques and their discordance in determination of the immune status in pregnancy women were not uncommon. IB test appeard to be highly useful in these situations as a confirmatory technique.Entities:
Keywords: Toxoplasma gondii; immunoblot; pregnant women; serodiagnosis; toxoplasmosis
Mesh:
Substances:
Year: 2013 PMID: 24039295 PMCID: PMC3770883 DOI: 10.3347/kjp.2013.51.4.485
Source DB: PubMed Journal: Korean J Parasitol ISSN: 0023-4001 Impact factor: 1.341
Results of EIA-IgG and FAT in the 2,136 tested sera
Results of LDBio-Toxo II IgG western blot with the 85 equivocal or discordant sera
Fig. 1Profiles with positive and negative blots in Toxo II IgG kit LDBIO®. PC: positive control with 5 bands: 30, 31, 33, 40, and 45 kDa; 1, positive blot with 3 bands (30, 31, and 33 kDa); 2, positive blot with 4 bands (30, 31, 33, and 40 kDa); 3, positive blot with 5 bands (30, 31, 33, 40, and 45 kDa); 4, negative blot without bands; 5, negative blot with 1 band (30 kDa); 6, negative blot showing 2 bands with low intensity (30 and 31 kDa).