Effect of high-pitch dual-source CT to compensate motion artifacts: a phantom study.

Nadja A Farshad-Amacker1, Hatem Alkadhi, Sebastian Leschka, Thomas Frauenfelder.   

Abstract

RATIONALE AND
OBJECTIVES: To evaluate the potential of high-pitch, dual-source computed tomography (DSCT) for compensation of motion artifacts.
MATERIALS AND METHODS: Motion artifacts were created using a moving chest/cardiac phantom with integrated stents at different velocities (from 0 to 4-6 cm/s) parallel (z direction), transverse (x direction), and diagonal (x and z direction combined) to the scanning direction using standard-pitch (SP) (pitch = 1) and high-pitch (HP) (pitch = 3.2) 128-detector DSCT (Siemens, Healthcare, Forchheim, Germany). The scanning parameters were (SP/HP): tube voltage, 120 kV/120 kV; effective tube current time product, 300 mAs/500 mAs; and a pitch of 1/3.2. Motion artifacts were analyzed in terms of subjective image quality and object distortion. Image quality was rated by two blinded, independent observers using a 4-point scoring system (1, excellent; 2, good with minor object distortion or blurring; 3, diagnostically partially not acceptable; and 4, diagnostically not acceptable image quality). Object distortion was assessed by the measured changes of the object's outer diameter (x) and length (z) and a corresponding calculated distortion vector (d) (d = √(x(2) + z(2))).
RESULTS: The interobserver agreement was excellent (k = 0.91). Image quality using SP was diagnostically not acceptable with any motion in x direction (scores 3 and 4), in contrast to HP DSCT where it remained diagnostic up to 2 cm/s (scores 1 and 2). For motion in the z direction only, image quality remained diagnostic for SP and HP DSCT (scores 1 and 2). Changes of the object's diameter (x), length (z), and distortion vectors (d) were significantly greater with SP (overall: x = 1.9 cm ± 1.7 cm, z = 0.6 cm ± 0.8 cm, and d = 1.4 cm ± 1.5 cm) compared to HP DSCT (overall: x = 0.1 cm ± 0.1 cm, z = 0.0 cm ± 0.1 cm, and d = 0.1 cm ± 0.1 cm; each P < .05).
CONCLUSION: High-pitch DSCT significantly decreases motion artifacts in various directions and improves image quality.
Copyright © 2013 AUR. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Phantom; high-pitch dual-source CT; motion artifacts

Mesh:

Year:  2013        PMID: 24029055     DOI: 10.1016/j.acra.2013.07.003

Source DB:  PubMed          Journal:  Acad Radiol        ISSN: 1076-6332            Impact factor:   3.173


  8 in total

1.  Coronary artery visibility in free-breathing young children on non-gated chest CT: impact of temporal resolution.

Authors:  Alexandre Bridoux; Antoine Hutt; Jean-Baptiste Faivre; Thomas Flohr; Alain Duhamel; Julien Pagniez; Jacques Remy; Martine Remy-Jardin
Journal:  Pediatr Radiol       Date:  2015-08-21

2.  Improved assessment of mediastinal and pulmonary pathologies in combined staging CT examinations using a fast-speed acquisition dual-source CT protocol.

Authors:  Franziska M Braun; Veronica Holzner; Felix G Meinel; Marco Armbruster; Martina Brandlhuber; Birgit Ertl-Wagner; Wieland H Sommer
Journal:  Eur Radiol       Date:  2017-07-04       Impact factor: 5.315

3.  Dual-source CT in blunt trauma patients: elimination of diaphragmatic motion using high-pitch spiral technique.

Authors:  Teresa Liang; Patrick McLaughlin; Chesnal D Arepalli; Luck J Louis; Ana-Maria Bilawich; John Mayo; Savvas Nicolaou
Journal:  Emerg Radiol       Date:  2015-12-04

4.  Performance of turbo high-pitch dual-source CT for coronary CT angiography: first ex vivo and patient experience.

Authors:  Fabian Morsbach; Sonja Gordic; Lotus Desbiolles; Daniela Husarik; Thomas Frauenfelder; Bernhard Schmidt; Thomas Allmendinger; Simon Wildermuth; Hatem Alkadhi; Sebastian Leschka
Journal:  Eur Radiol       Date:  2014-05-17       Impact factor: 5.315

5.  Influence of Heart Rate and Innovative Motion-Correction Algorithm on Coronary Artery Image Quality and Measurement Accuracy Using 256-Detector Row Computed Tomography Scanner: Phantom Study.

Authors:  Jeong Bin Park; Yeon Joo Jeong; Geewon Lee; Nam Kyung Lee; Jin You Kim; Ji Won Lee
Journal:  Korean J Radiol       Date:  2018-12-27       Impact factor: 3.500

6.  Simultaneous Assessment of Left Ventricular Function and Coronary Artery Anatomy by Third-generation Dual-source Computed Tomography Using a Low Radiation Dose.

Authors:  Ji Won Lee; Kyung Jin Nam; Jin You Kim; Yeon Joo Jeong; Geewon Lee; So Min Park; Soo Jin Lim; Ki Seok Choo
Journal:  J Cardiovasc Imaging       Date:  2019-11-20

7.  Evidence-based recommendations for musculoskeletal kinematic 4D-CT studies using wide area-detector scanners: a phantom study with cadaveric correlation.

Authors:  Pedro Augusto Gondim Teixeira; Anne-Sophie Formery; Gabriela Hossu; Daniel Winninger; Toufik Batch; Alban Gervaise; Alain Blum
Journal:  Eur Radiol       Date:  2016-04-19       Impact factor: 5.315

8.  Image Quality and Radiation Dose of High-Pitch Dual-Source Spiral Cardiothoracic Computed Tomography in Young Children with Congenital Heart Disease: Comparison of Non-Electrocardiography Synchronization and Prospective Electrocardiography Triggering.

Authors:  Hyun Woo Goo
Journal:  Korean J Radiol       Date:  2018-10-18       Impact factor: 3.500

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.