Literature DB >> 24021927

Quantification of ventilation distribution in regional lung injury by electrical impedance tomography and xenon computed tomography.

Gunnar Elke1, Matthew K Fuld, Ahmed F Halaweish, Bartłomiej Grychtol, Norbert Weiler, Eric A Hoffman, Inéz Frerichs.   

Abstract

Validation studies of electrical impedance tomography (EIT) based assessment of regional ventilation under pathological conditions are required to prove that EIT can reliably quantify heterogeneous ventilation distribution with sufficient accuracy. The objective of our study was to validate EIT measurements of regional ventilation through a comparison with xenon-multidetector-row computed tomography (XeCT) in an animal model of sub-lobar lung injury. Nine anesthetized mechanically ventilated supine pigs were examined before and after the induction of lung injury in two adjacent sub-lobar segments of the right lung by saline lavage or endotoxin instillation. Regional ventilation was determined in 32 anteroposterior regions of interest in the right and left lungs and the ventilation change quantified by difference images between injury and control. Six animals were included in the final analysis. Measurements of regional ventilation by EIT and XeCT correlated well before (rs = 0.89 right, rs = 0.90 left lung) and after local injury (rs = 0.79 and 0.92, respectively). No bias and narrow limits of agreement were found during both conditions. The ventilation decrease in the right injured lung was correspondingly measured by both modalities (5.5%±1.1% by EIT and 5.4%±1.9% by XeCT, p = 0.94). EIT was inferior to clearly separate the exact anatomical location of the regional injuries. Regional ventilation was overestimated (<2%) in the most ventral and dorsal regions and underestimated (2%) in the middle regions by EIT compared to XeCT. This study shows that EIT is able to reliably discern even small ventilation changes on sub-lobar level.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24021927     DOI: 10.1088/0967-3334/34/10/1303

Source DB:  PubMed          Journal:  Physiol Meas        ISSN: 0967-3334            Impact factor:   2.833


  6 in total

1.  Chest electrical impedance tomography examination, data analysis, terminology, clinical use and recommendations: consensus statement of the TRanslational EIT developmeNt stuDy group.

Authors:  Inéz Frerichs; Marcelo B P Amato; Anton H van Kaam; David G Tingay; Zhanqi Zhao; Bartłomiej Grychtol; Marc Bodenstein; Hervé Gagnon; Stephan H Böhm; Eckhard Teschner; Ola Stenqvist; Tommaso Mauri; Vinicius Torsani; Luigi Camporota; Andreas Schibler; Gerhard K Wolf; Diederik Gommers; Steffen Leonhardt; Andy Adler
Journal:  Thorax       Date:  2016-09-05       Impact factor: 9.139

2.  EIT monitors valid and robust regional ventilation distribution in pathologic ventilation states in porcine study using differential DualEnergy-CT (ΔDECT).

Authors:  Sebastian D Reinartz; Michael Imhoff; René Tolba; Felix Fischer; Eike G Fischer; Eckhard Teschner; Sabine Koch; Yvo Gärber; Peter Isfort; Felix Gremse
Journal:  Sci Rep       Date:  2019-07-05       Impact factor: 4.379

3.  Perioperative redistribution of regional ventilation and pulmonary function: a prospective observational study in two cohorts of patients at risk for postoperative pulmonary complications.

Authors:  Maria Bauer; Anne Opitz; Jörg Filser; Hendrik Jansen; Rainer H Meffert; Christoph T Germer; Norbert Roewer; Ralf M Muellenbach; Markus Kredel
Journal:  BMC Anesthesiol       Date:  2019-07-27       Impact factor: 2.217

4.  Effect of Electrode Belt and Body Positions on Regional Pulmonary Ventilation- and Perfusion-Related Impedance Changes Measured by Electric Impedance Tomography.

Authors:  Elin Ericsson; Erik Tesselaar; Folke Sjöberg
Journal:  PLoS One       Date:  2016-06-02       Impact factor: 3.240

5.  Intratracheal exposure of common marmosets to MERS-CoV Jordan-n3/2012 or MERS-CoV EMC/2012 isolates does not result in lethal disease.

Authors:  Reed F Johnson; Laura E Via; Mia R Kumar; Joseph P Cornish; Srikanth Yellayi; Louis Huzella; Elena Postnikova; Nicholas Oberlander; Christopher Bartos; Britini L Ork; Steven Mazur; Cindy Allan; Michael R Holbrook; Jeffrey Solomon; Joshua C Johnson; James Pickel; Lisa E Hensley; Peter B Jahrling
Journal:  Virology       Date:  2015-09-03       Impact factor: 3.616

6.  Effects of individualized electrical impedance tomography and image reconstruction settings upon the assessment of regional ventilation distribution: Comparison to 4-dimensional computed tomography in a porcine model.

Authors:  Florian Thürk; Stefan Boehme; Daniel Mudrak; Stefan Kampusch; Alice Wielandner; Helmut Prosch; Christina Braun; Frédéric P R Toemboel; Johannes Hofmanninger; Eugenijus Kaniusas
Journal:  PLoS One       Date:  2017-08-01       Impact factor: 3.240