H K Maradiya1, Vasundhara H Pansara.
Abstract
A simple, accurate, precise and sensitive reverse-phase high-performance liquid chromatography method for the determination of citicoline sodium has been developed and validated. Drug was resolved on a C18 column (Phenomenex Luna, 250×4.6 mm, 10μ), utilizing mobile phase of potassium dihydrogen phosphate buffer and acetonitrile in a ratio of 30:70. Mobile phase was delivered at the flow rate of 1.0 ml/min and detection was carried out at 272 nm. Separation was completed within 2.22 min. Calibration curve was linear with good correlation coefficient (R(2)=0.999) over a concentration range 10-60 μg/ml. Recovery was between 98.84 and 101.76%. Method was found to be reproducible with relative standard deviation for intra and interday precision of <2.0% over the said concentration range. The method was successfully applied to the determination of the citicoline sodium, it can be very useful and an alternate to performing the stability studies.Entities:
Keywords: Acetonitrile; citicoline sodium; high performance liquid chromatography analysis; potassium dihydrogen phosphate
Year: 2013 PMID: 24019577 PMCID: PMC3757867
Source DB: PubMed Journal: Indian J Pharm Sci ISSN: 0250-474X Impact factor: 0.975
Fig. 1CONDITION USED FOR CHROMATOGRAPHY ANALYSIS
SUMMARY OF VALIDATION AND SST PARAMETERS
Fig. 2HPLC chromatogram.
high-performance liquid chromatogram for the analysis of citicoline sodium reference standard by developed validated method.
RESULTS FOR RECOVERY STUDY
DETERMINATION OF CITICOLINE SODIUM IN TABLETS AND COMPARISON WITH REFERENCE METHOD