Simultaneous anterior and posterior compression of the pelvic ring with external fixation using a pre-tensed curved bar: A biomechanical study.

Alfonso Queipo-de-Llano1, Ana Perez-Blanca, Francisco Ezquerro, Felipe Luna-González.   

Abstract

INTRODUCTION: External fixators continue to be essential tools in the urgent treatment of pelvic fractures for compression and stabilization of the pelvic ring. Current systems fail to produce simultaneous anterior and posterior compression. A modified application of an existing curved bar fixator is proposed using a specifically designed tensioner to pre-tense the bar prior to its connection to Schanz screws. Subsequent pre-tension release and elastic recovery of the bar could potentially compress the pelvis. The aim of this work was to determine if the modified application could produce greater simultaneous compression across the sacroiliac joint and the symphysis of an unstable fractured pelvis than the standard application without pretension.
MATERIALS AND METHODS: Six synthetic pelvis models with symphyseal and unilateral sacroiliac joints disruptions, simulating a Tile type C pelvic ring fracture, were used. Each specimen was stabilized using two 5mm×250mm supra-acetabular Schanz pins, a couple of open adjustable clamps and a semicircular carbon fibre rod applied without and with pre-tension. Two distances from bar to bone and three levels of pretension were compared. Each pelvis was tested with the six possible parameter combinations. Compressive forces at the disrupted joints were measured using pressure sensitive film sensors.
RESULTS: The modified application produced forces significantly higher than the minimal compression achieved with standard application. At the sacroiliac joint, after pre-tension release, mean compressive forces measured ranged from 28.7 to 85.6N. The closest bar-to-bone distance always produced a significantly higher force; similarly, a significant increase in compression was found as the pre-tension level rose. At the symphysis, mean compressive forces between 35.3N and 49.0N were determined. No significant variations were seen with changes of any of the two factors analyzed
CONCLUSIONS: To pre-tense a semi-circular bar before its use for external fixation of the fractured pelvis, is an effective means of applying compression simultaneously through the sacroiliac joint and the symphysis. The proposed method generates the highest compressive forces at the sacroiliac joint when the rod is subject to the highest pre-tension level not producing subluxation and is subsequently positioned as close as possible to the bone depending on patient's condition.
Copyright © 2013 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Biomechanical study; External fixation; Pelvic compression; Pelvis fracture; Pre-tensed bar; Urgent stabilization

Mesh:

Year:  2013        PMID: 24016457     DOI: 10.1016/j.injury.2013.08.016

Source DB:  PubMed          Journal:  Injury        ISSN: 0020-1383            Impact factor:   2.586


  4 in total

1.  Achievable pin spanning angulation in anterosuperior pelvic external fixation.

Authors:  Benjamin Young; Drayton Daily; Clarence Kee; Kevin Perry; Massimo Max Morandi; R Shane Barton; Giovanni F Solitro
Journal:  Eur J Orthop Surg Traumatol       Date:  2020-05-01

2.  Resuscitation of Polytrauma Patients: The Management of Massive Skeletal Bleeding.

Authors:  Enrique Guerado; Maria Luisa Bertrand; Luis Valdes; Encarnacion Cruz; Juan Ramon Cano
Journal:  Open Orthop J       Date:  2015-07-31

3.  Early Experience with Reduction of Unstable Pelvic Fracture Using a Computer-Aided Reduction Frame.

Authors:  Jing-Xin Zhao; Li-Cheng Zhang; Xiu-Yun Su; Zhe Zhao; Yan-Peng Zhao; Guo-Fei Sun; Li-Hai Zhang; Pei-Fu Tang
Journal:  Biomed Res Int       Date:  2018-02-13       Impact factor: 3.411

4.  Sacrospinous and sacrotuberous ligaments influence in pelvis kinematics.

Authors:  Petr Henyš; Maziar Ramezani; Daniel Schewitz; Andreas Höch; Dustin Möbius; Benjamin Ondruschka; Niels Hammer
Journal:  J Anat       Date:  2022-08-20       Impact factor: 2.921

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.