Abstract
A sensitive and simple method for the methotrexate quantification was developed using aminopterin as internal standard. Methotrexate is an anticancer agent that is widely used in a variety of human cancers including primary central nervous system lymphoma. The compound was quantified by liquid-chromatography coupled to electrospray ionization (positive ion-mode) low-energy collision dissociation-tandem mass spectrometry. Quantitative detection was by multiple reaction monitoring of the transitions of the [M+H]+ ion of MTX to its common product ion at m/z 308.4 and of aminopterin at m/z 441.2→m/z 294.0. The method demonstrated linearity over at least three orders of magnitude and had a detection limit of 1ng/ml for methotrexate. A run time of less than 8.0min for each sample made it possible to analyze a large number of human saliva samples per day. Application of this procedure was demonstrated to a saliva excretion study of methotrexate on the samples obtained after an intravenously administration of 1mg/kg/dose of methotrexate to six patients with acute lymphoblastic leukemia.Entities:
Keywords: Liquid chromatography; Methotrexate; Saliva excretion study; Tandem mass spectrometry
Mesh:
Substances:
Year: 2013 PMID: 24008118 DOI: 10.1016/j.jchromb.2013.07.026