Central-third bone-patellar tendon-bone allografts demonstrate superior biomechanical failure characteristics compared with hemi-patellar tendon grafts.

Adam B Yanke1, Rebecca Bell, Andrew S Lee, Elizabeth Shewman, Vincent M Wang, Bernard R Bach.   

Abstract

BACKGROUND: Reconstruction of the anterior cruciate ligament (ACL) is commonly performed with a bone-patellar tendon-bone (BTB) allograft. However, grafts may result from harvesting the central region of a whole graft (C-BTB), the medial 10 mm of a lateral hemi-BTB (L-BTB) graft, or the lateral 10 mm of a medial hemi-BTB (M-BTB) graft.
PURPOSE: To quantify potential differences in graft biomechanical properties when comparing whole versus hemi-BTB grafts. STUDY
DESIGN: Controlled laboratory study.
METHODS: Ten pairs of human BTB allografts (irradiated with 1.0-1.2 Mrad) were randomized to preparation as whole grafts or hemigrafts. From these, 10-mm grafts were prepared from the center or the most central portion, respectively. After measurements of tendon thickness, width, and length, specimens underwent cyclic tensile testing, followed by load-to-failure analysis. Biomechanical outcomes included cyclic elongation and creep strain along with the following failure characteristics: maximum load, elongation at maximum load, maximum stress, strain at maximum stress, and linear stiffness.
RESULTS: Regionally, the mean thickness of the C-BTB (5.18 ± 0.75 mm), M-BTB (5.08 ± 0.56 mm), and L-BTB (5.32 ± 0.62 mm) grafts were comparable (P > .72). Similarly, the mean length of the C-BTB (47.4 ± 6.73 mm), M-BTB (47.0 ± 5.45 mm), and L-BTB (50.7 ± 6.42 mm) grafts were alike (P > .43). While differences in cyclic elongation and strain were not significant, the M-BTB graft tended to elongate more (0.204 ± 0.13 mm; P = .075) and experience greater strain (0.56% ± 0.32%; P = .054) compared with the C-BTB graft (0.09 ± 0.03 mm and 0.23% ± 0.07%, respectively). Load-to-failure testing demonstrated a higher maximum load (2293 ± 531 N) and stiffness (356 ± 46 N/mm) of the C-BTB graft as compared with the M-BTB graft (1575 ± 325 N [P < .007] and 275 ± 37 N/mm [P < .008], respectively) and L-BTB graft (1585 ± 452 N [P < .008] and 277 ± 65 N/mm [P < .009], respectively). No differences were noted with respect to elongation or stress at maximum load among the grafts. Maximum stress in the C-BTB graft (45.4 ± 11.5 MPa) was greater than in the L-BTB graft (29.7 ± 10.6 MPa) (P < .03) and tended to be greater than the M-BTB graft (34.1 ± 6.27 MPa) (P = .087).
CONCLUSION: Biomechanical failure properties (maximum load, stress, and stiffness) of the central portion of a whole BTB graft are superior to those of the medial portion of a lateral hemi-BTB graft and the lateral portion of a medial hemi-BTB graft. However, cyclic loading characteristics did not differ between grafts. CLINICAL RELEVANCE: Although the true central-third BTB graft is biomechanically superior to hemi-BTB grafts, future studies are necessary to determine if the use of hemigrafts leads to an increased incidence of clinical failure.

Entities:  

Keywords:  ACL; allografts; anterior cruciate ligament; biomechanics of ligament; bone–patellar tendon–bone allograft; knee; ligaments; morphology

Mesh:

Year:  2013        PMID: 24007760     DOI: 10.1177/0363546513501780

Source DB:  PubMed          Journal:  Am J Sports Med        ISSN: 0363-5465            Impact factor:   6.202


  4 in total

1.  Can the gracilis replace the anterior cruciate ligament in the knee? A biomechanical study.

Authors:  Etienne Cavaignac; Regis Pailhé; Nicolas Reina; Jérôme Murgier; Jean Michel Laffosse; Philippe Chiron; Pascal Swider
Journal:  Int Orthop       Date:  2015-11-05       Impact factor: 3.075

Review 2.  What Factors Influence the Biomechanical Properties of Allograft Tissue for ACL Reconstruction? A Systematic Review.

Authors:  Drew A Lansdown; Andrew J Riff; Molly Meadows; Adam B Yanke; Bernard R Bach
Journal:  Clin Orthop Relat Res       Date:  2017-10       Impact factor: 4.176

3.  Allograft tissue irradiation and failure rate after anterior cruciate ligament reconstruction: A systematic review.

Authors:  Jesse Dashe; Robert L Parisien; Antonio Cusano; Emily J Curry; Asheesh Bedi; Xinning Li
Journal:  World J Orthop       Date:  2016-06-18

4.  Our Preferred Technique for Bone-Patellar Tendon-Bone Allograft Preparation.

Authors:  Steven DeFroda; Michael Fice; Sarah Tepper; Bernard R Bach
Journal:  Arthrosc Tech       Date:  2021-11-22
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.