Octavius 4D characterization for flattened and flattening filter free rotational deliveries.

Conor K McGarry1, Barry F O'Connell, Mark W D Grattan, Christina E Agnew, Denise M Irvine, Alan R Hounsell.   

Abstract

PURPOSE: In this study the Octavius detector 729 ionization chamber (IC) array with the Octavius 4D phantom was characterized for flattening filter (FF) and flattening filter free (FFF) static and rotational beams. The device was assessed for verification with FF and FFF RapidArc treatment plans.
METHODS: The response of the detectors to field size, dose linearity, and dose rate were assessed for 6 MV FF beams and also 6 and 10 MV FFF beams. Dosimetric and mechanical accuracy of the detector array within the Octavius 4D rotational phantom was evaluated against measurements made using semiflex and pinpoint ionization chambers, and radiochromic film. Verification FF and FFF RapidArc plans were assessed using a gamma function with 3%∕3 mm tolerances and 2%∕2 mm tolerances and further analysis of these plans was undertaken using film and a second detector array with higher spatial resolution.
RESULTS: A warm-up dose of >6 Gy was required for detector stability. Dose-rate measurements were stable across a range from 0.26 to 15 Gy∕min and dose response was linear, although the device overestimated small doses compared with pinpoint ionization chamber measurements. Output factors agreed with ionization chamber measurements to within 0.6% for square fields of side between 3 and 25 cm and within 1.2% for 2 × 2 cm(2) fields. The Octavius 4D phantom was found to be consistent with measurements made with radiochromic film, where the gantry angle was found to be within 0.4° of that expected during rotational deliveries. RapidArc FF and FFF beams were found to have an accuracy of >97.9% and >90% of pixels passing 3%∕3 mm and 2%∕2 mm, respectively. Detector spatial resolution was observed to be a factor in determining the accurate delivery of each plan, particularly at steep dose gradients. This was confirmed using data from a second detector array with higher spatial resolution and with radiochromic film.
CONCLUSIONS: The Octavius 4D phantom with associated Octavius detector 729 ionization chamber array is a dosimetrically and mechanically stable device for pretreatment verification of FF and FFF RapidArc treatments. Further improvements may be possible through use of a detector array with higher spatial resolution (detector size and∕or detector spacing).

Entities:  

Mesh:

Year:  2013        PMID: 24007140     DOI: 10.1118/1.4817482

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  8 in total

1.  A study on the correlation between plan complexity and gamma index analysis in patient specific quality assurance of volumetric modulated arc therapy.

Authors:  Dhanabalan Rajasekaran; Prakash Jeevanandam; Prabakar Sukumar; Arulpandiyan Ranganathan; Samdevakumar Johnjothi; Vivekanandan Nagarajan
Journal:  Rep Pract Oncol Radiother       Date:  2014-09-06

2.  Treatment of breast cancer with simultaneous integrated boost in hybrid plan technique : Influence of flattening filter-free beams.

Authors:  Marzieh Bahrainy; Matthias Kretschmer; Vincent Jöst; Astrid Kasch; Florian Würschmidt; Jörg Dahle; Jörn Lorenzen
Journal:  Strahlenther Onkol       Date:  2016-03-14       Impact factor: 3.621

3.  Class solutions for SABR-VMAT for high-risk prostate cancer with and without elective nodal irradiation.

Authors:  Sarah O S Osman; Prakash Jeevanandam; Nithya Kanakavelu; Denise M Irvine; Ciara A Lyons; Suneil Jain; Alan R Hounsell; Conor K McGarry
Journal:  Radiat Oncol       Date:  2016-11-24       Impact factor: 3.481

4.  Evaluation of semiempirical VMAT dose reconstruction on a patient dataset based on biplanar diode array measurements.

Authors:  Cassandra Stambaugh; Daniel Opp; Stuart Wasserman; Geoffrey Zhang; Vladimir Feygelman
Journal:  J Appl Clin Med Phys       Date:  2014-03-06       Impact factor: 2.102

5.  Verification of high-dose-rate brachytherapy treatment planning dose distribution using liquid-filled ionization chamber array.

Authors:  A B Mohamed Yoosuf; Prakash Jeevanandam; Glenn Whitten; Geraldine Workman; Conor K McGarry
Journal:  J Contemp Brachytherapy       Date:  2018-04-30

6.  Volumetric-modulated Arc Therapy Lung Stereotactic Body Radiation Therapy Dosimetric Quality Assurance: A Comparison between Radiochromic Film and Chamber Array.

Authors:  Juan Fernando Mata Colodro; Alfredo Serna Berná; Vicente Puchades Puchades; David Ramos Amores; Miguel Alcaraz Baños
Journal:  J Med Phys       Date:  2017 Jul-Sep

7.  Correlation of phantom-based and log file patient-specific QA with complexity scores for VMAT.

Authors:  Christina E Agnew; Denise M Irvine; Conor K McGarry
Journal:  J Appl Clin Med Phys       Date:  2014-11-08       Impact factor: 2.102

8.  Practical application of Octavius® -4D: Characteristics and criticalities for IMRT and VMAT verification.

Authors:  Patrizia Urso; Rita Lorusso; Luca Marzoli; Daniela Corletto; Paolo Imperiale; Annalisa Pepe; Lorenzo Bianchi
Journal:  J Appl Clin Med Phys       Date:  2018-07-16       Impact factor: 2.102

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.