Abstract
Sensitive and selective analytical procedures based on high-performance liquid chromatography with mass spectrometric detection were developed for the determination of amoxicillin in human plasma samples. Samples were prepared by applying in-house manufactured molecularly imprinted solid-phase microextraction probes. The detection of target compounds was performed in multiple reaction monitoring mode. The multiple reaction monitoring detection was operated in the positive electrospray ionization mode using the transitions of m/z 366 ([M + H](+) ) → 349 for amoxicillin and m/z 390 ([M + H](+) ) → 372 for gemifloxacin. The method was validated with precision within 15% relative standard deviation and accuracy within 15% relative error. The method was successfully applied to study of the pharmacokinetics of amoxicillin in human plasma after oral administration of amoxicillin.Entities:
Keywords: antibiotic drugs; high performance liquid chromatography; mass spectrometry; sample preparation techniques
Mesh:
Substances:
Year: 2013 PMID: 24006329 DOI: 10.1002/bmc.3014
Source DB: PubMed Journal: Biomed Chromatogr ISSN: 0269-3879 Impact factor: 1.902