Validation of an MRI-based method to assess patellofemoral joint contact areas in loaded knee flexion in vivo.

Emily J McWalter1, Colm M O'Kane, David P Fitzpatrick, David R Wilson.   

Abstract

PURPOSE: To develop and validate short axial and sagittal MRI scans (<1min) to assess in vivo patellofemoral contact areas in loaded knee flexion.
MATERIALS AND METHODS: Contact area was assessed in four cadaver knee specimens from axial and sagittal scans using two contact area extraction techniques (delineation and intersection) and three calculation techniques (slice thickness multiplication, linear interpolation, and spline interpolation). Error was expressed as the mean absolute and percentage difference from a dye staining-based reference standard. Intrareader and intrasubject repeatability, expressed as the mean standard deviation, was determined.
RESULTS: Contact area assessments from the sagittal MRI scans using the delineation and slice thickness multiplication technique had the smallest error (47.7 ± 38.1 mm(2) or 10.7%). The intrareader repeatability from assessments using the sagittal scans was smaller than those using the axial scans when the delineation method was used (<9.4 ± 4.3 mm(2) and <15.4 ± 14.1 mm(2) , respectively). The intrasubject repeatability of the assessment from the sagittal scan was less than 39.9 ± 23.0 mm(2) .
CONCLUSION: This protocol yields assessments of contact area in less than 1 minute that have errors similar to those made using scans many times longer and can be used in series with kinematic scans to carry out simultaneous assessments in vivo to study patellofemoral joint disease.
Copyright © 2013 Wiley Periodicals, Inc.

Entities:  

Keywords:  contact area; kinematics; knee; magnetic resonance imaging; patella; patellofemoral; validation

Mesh:

Year:  2013        PMID: 24006182     DOI: 10.1002/jmri.24240

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  4 in total

1.  Inter- and intra-rater reliability of patellofemoral kinematic and contact area quantification by fast spin echo MRI and correlation with cartilage health by quantitative T1ρ MRI.

Authors:  Brian C Lau; Daniel U Thuillier; Valentina Pedoia; Ellison Y Chen; Zhihong Zhang; Brian T Feeley; Richard B Souza
Journal:  Knee       Date:  2015-12-31       Impact factor: 2.199

2.  Validation of a method for combining biplanar radiography and magnetic resonance imaging to estimate knee cartilage contact.

Authors:  Eric Thorhauer; Scott Tashman
Journal:  Med Eng Phys       Date:  2015-08-21       Impact factor: 2.242

Review 3.  Dynamic MRI for articulating joint evaluation on 1.5 T and 3.0 T scanners: setup, protocols, and real-time sequences.

Authors:  Marc Garetier; Bhushan Borotikar; Karim Makki; Sylvain Brochard; François Rousseau; Douraïed Ben Salem
Journal:  Insights Imaging       Date:  2020-05-19

4.  Reliability of tibiofemoral contact area and centroid location in upright, open MRI.

Authors:  Andrew M Schmidt; David J Stockton; Michael A Hunt; Andrew Yung; Bassam A Masri; David R Wilson
Journal:  BMC Musculoskelet Disord       Date:  2020-11-30       Impact factor: 2.362

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.