Literature DB >> 24004981

Influence of beach chair position on cerebral oxygen saturation: a comparison of INVOS and FORE-SIGHT cerebral oximeter.

Dorothea Closhen1, Manfred Berres, Christian Werner, Kristin Engelhard, Patrick Schramm.   

Abstract

BACKGROUND: Although beach chair position (BCP) is frequently used for shoulder surgery, a potentially detrimental influence on cerebral oxygenation is discussed. Therefore, the present study investigated changes in regional cerebral oxygen saturation (rSO2/StO2) during BCP comparing 2 different devices for near-infrared spectroscopy measurement.
METHODS: Data were collected in 35 patients undergoing shoulder surgery in BCP and compared with a control group of 35 awake volunteers. The rSO2/StO2 was assessed using INVOS and FORE-SIGHT monitors. Mean arterial blood pressure (MAP), peripheral oxygen saturation (SpO2), PeCO2, FiO2, end-tidal sevoflurane concentration, and rSO2/StO2 were measured before positioning, during BCP, and in supine position after surgery.
RESULTS: A decrease in rSO2/StO2 could be observed after BCP (INVOS: 76.1% supine vs. 66.7% BCP, P<0.001; FORE-SIGHT: 78.6% supine, 66.1% BCP, P<0.001), which was reversible in supine position. This decrease correlated with MAP during BCP, while in supine position no correlation was detected. In control group BCP did not influence rSO2/StO2. Changes detected with INVOS or FORE-SIGHT cerebral oximeter did not differ.
CONCLUSION: BCP is associated with a decrease in rSO2/StO2 of 10% in anesthetized patients, which is reversible after repositioning. No changes occurred in supine position under general anesthesia as well as in awake subjects in BCP. This underlines the assumption that vasodilation by anesthetics in combination with BCP evoke a drop in rSO2/StO2. A strict hemodynamic management may be necessary to prevent desaturation events. Despite different technology used by the devices, the results of INVOS and FORE-SIGHT cerebral oximeters are comparable.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24004981     DOI: 10.1097/ANA.0b013e3182973349

Source DB:  PubMed          Journal:  J Neurosurg Anesthesiol        ISSN: 0898-4921            Impact factor:   3.956


  7 in total

1.  Effect of posture and extracranial contamination on results of cerebral oximetry by near-infrared spectroscopy.

Authors:  Shinya Kato; Kenji Yoshitani; Yosuke Kubota; Yuzuru Inatomi; Yoshihiko Ohnishi
Journal:  J Anesth       Date:  2016-11-02       Impact factor: 2.078

Review 2.  Cerebral near-infrared spectroscopy (NIRS) for perioperative monitoring of brain oxygenation in children and adults.

Authors:  Yun Yu; Kaiying Zhang; Ling Zhang; Huantao Zong; Lingzhong Meng; Ruquan Han
Journal:  Cochrane Database Syst Rev       Date:  2018-01-17

3.  Detection of critical cerebral desaturation thresholds by three regional oximeters during hypoxia: a pilot study in healthy volunteers.

Authors:  Kerry L Tomlin; Anna-Maria Neitenbach; Ulf Borg
Journal:  BMC Anesthesiol       Date:  2017-01-13       Impact factor: 2.217

Review 4.  Cerebral Desaturation Events During Shoulder Arthroscopy in the Beach Chair Position.

Authors:  Dane H Salazar; William J Davis; Nezih Ziroğlu; Nickolas G Garbis
Journal:  J Am Acad Orthop Surg Glob Res Rev       Date:  2019-08-02

5.  Cerebral oximetry-guided pulmonary artery banding for end-stage heart failure in a child with left ventricular noncompaction cardiomyopathy: a case report.

Authors:  Mayu Asano; Kenji Doi; Minoru Nomura; Yasuko Nagasaka
Journal:  Transl Pediatr       Date:  2021-11

Review 6.  Near-infrared spectroscopy for the evaluation of anesthetic depth.

Authors:  Gabriela Hernandez-Meza; Meltem Izzetoglu; Mary Osbakken; Michael Green; Kurtulus Izzetoglu
Journal:  Biomed Res Int       Date:  2015-10-01       Impact factor: 3.411

7.  Adverse heart rate responses during beach-chair position for shoulder surgeries - A systematic review and meta-analysis of their incidence, interpretations and associations.

Authors:  Thrivikrama Padur Tantry; Harish Karanth; Reshma Koteshwar; Pramal K Shetty; Karunakara K Adappa; Sunil P Shenoy; Dinesh Kadam; Sudarshan Bhandary
Journal:  Indian J Anaesth       Date:  2020-07-31
  7 in total