Husniza Hussain1, Norhayati Mustafa Khalid, Rusidah Selamat, Wan Mohamud Wan Nazaimoon.
Abstract
BACKGROUND: The urinary iodine micromethod (UIMM) is a modification of the conventional method and its performance needs evaluation.Entities:
Keywords: Method acceptability; Method evaluation; Quality control; Sigma metrics; Urinary iodine
Mesh:
Substances:
Year: 2013 PMID: 24003421 PMCID: PMC3756235 DOI: 10.3343/alm.2013.33.5.319
Source DB: PubMed Journal: Ann Lab Med ISSN: 2234-3806 Impact factor: 3.464
Total allowable errors (TEas) set by two urinary iodine (UI) external quality assurance (EQA) providers
Abbreviations: TUIQP, The Urinary Iodine Quality Program, Institute for Clinical Pathology and Medical Research, Australia; EQUIP, Ensuring the Quality of Urinary Iodine Procedures program, Centre for Disease Control, USA.
The method acceptability tests results on standard deviations and errors from the method validation data with criteria check against two EQA providers total allowable errors (TEas)
Abbreviations: S w-run, within-run standard deviation; S tot, total standard deviation; SE, systematic error; RE, random error; TEcalc, calculated total error; z, multiplier representing desired confidence interval; TUIQP, The Urinary Iodine Quality Program, Institute for Clinical Pathology and Medical Research, Australia; EQUIP, Ensuring the Quality of Urinary Iodine Procedures program, Centre for Disease Control, USA.
The method acceptability tests results on t-value and method comparison study from the method validation data
Statistical formula definition are as stated in the reference [13].
Abbreviations: SE, systematic error; RE, random error; UIMM, urinary iodine micromethod.
Reportable range worksheet for quantifying errors from the method validation linearity study
Abbreviations: TUIQP, The Urinary Iodine Quality Program, Institute for Clinical Pathology and Medical Research, Australia; EQUIP, Ensuring the Quality of Urinary Iodine Procedures program, Centre for Disease Control, USA.
Fig. 1Urinary iodine micromethod's normalized method decision chart against The Urinary Iodine Quality Program (TUIQP) TEas based on the 2008 National IDD Survey EQA results (Lab A [n=20], Lab B [n=18], Lab C [n=12], and Lab D [n=6]) and the 2006 method validation study (MV). According to urinary iodine ranges in Table 1.
Abbreviations: L, low control; M-L, medium-low control; M-H, medium-high control; H, high control.
Systematic error (SE), random error (RE), and calculated total error (TEcalc) for the urinary iodine micromethod performed by the four IDD laboratories during the 2008 National IDD Survey (Lab A [n=20], Lab B [n=18], Lab C [n=12], and Lab D [n=6])
Abbreviations: TUIQP, The Urinary Iodine Quality Program, Institute for Clinical Pathology and Medical Research, Australia; EQUIP, Ensuring the Quality of Urinary Iodine Procedures program, Centre for Disease Control, USA.
Fig. 2Urinary iodine micromethod's normalized method decision chart based on the 2008 national IDD survey EQA results (Lab A [n=20], Lab B [n=18], Lab C [n=12], and Lab D [n=6]) and against EQUIP TEas. According to urinary iodine ranges in Table 1.
Abbreviations: L, low control; M-H, medium-high control; H, high control.