Literature DB >> 24002030

Cardiac perforation from implantable cardioverter-defibrillator lead placement: insights from the national cardiovascular data registry.

Jonathan C Hsu1, Paul D Varosy, Haikun Bao, Thomas A Dewland, Jeptha P Curtis, Gregory M Marcus.   

Abstract

BACKGROUND: Cardiac perforation is a feared complication of implantable cardioverter-defibrillator (ICD) lead implantation because of the potential for significant morbidity and mortality. Predictors of perforation and the severity of associated adverse events have not been well studied. We sought to identify predictors of cardiac perforation from ICD lead implantation and subsequent outcomes. METHODS AND
RESULTS: We studied 440 251 first-time ICD recipients in the ICD Registry implanted between January 2006 and September 2011. Using hierarchical multivariable logistic regression adjusting for patient, implanting physician, and hospital characteristics, we examined the predictors of perforation and the association of perforation with other major complications, length of stay, and in-hospital mortality. Cardiac perforation occurred in 625 patients (0.14%). After multivariable adjustment, older age, female sex, left bundle branch block, worsened heart failure class, higher left ventricular ejection fraction, and non-single-chamber ICD implant were associated with a greater odds of perforation. Conversely, atrial fibrillation, diabetes mellitus, previous cardiac bypass surgery, and higher implanter procedural volume were associated with a lower odds of perforation (all P<0.05). After adjustment, ICD recipients with perforation had greater odds of other associated major complications (odds ratio, 27.5; 95% confidence interval, 19.9-38.0; P<0.0001), postprocedural hospital stays >3 days (odds ratio, 16.3; 95% confidence interval, 13.7-19.4; P<0.0001), and in-hospital death (odds ratio, 17.7; 95% confidence interval, 12.2-25.6; P<0.0001).
CONCLUSIONS: In a large population of ICD recipients, specific patient and implanter characteristics predicted cardiac perforation risk. Cardiac perforation was associated with a substantially increased risk of other major complications, prolonged hospital stays, and death.

Entities:  

Keywords:  adverse events complication; cardiac perforation; complication; defibrillators, implantable; mortality; national registry

Mesh:

Year:  2013        PMID: 24002030     DOI: 10.1161/CIRCOUTCOMES.113.000299

Source DB:  PubMed          Journal:  Circ Cardiovasc Qual Outcomes        ISSN: 1941-7713


  10 in total

1.  Pleural Effusion Secondary to Atrial Perforation During COVID-19 Period.

Authors:  Eduardo Laviña Soriano; Yunelsy Anta Mejías; Lucía Gil Abadía; Olga Mediano
Journal:  Arch Bronconeumol       Date:  2020-10-13       Impact factor: 4.872

2.  Utilization and in-hospital complications of cardiac resynchronization therapy: trends in the United States from 2003 to 2013.

Authors:  Seyed Mohammadreza Hosseini; Kasra Moazzami; Guy Rozen; Jeena Vaid; Ahmed Saleh; Kevin E Heist; Mark Vangel; Jeremy N Ruskin
Journal:  Eur Heart J       Date:  2017-07-14       Impact factor: 29.983

3.  Serious case of the hiccups.

Authors:  Elizabeth V Saarel; Kevin Hinkle; Susan P Etheridge
Journal:  HeartRhythm Case Rep       Date:  2015-02-13

4.  Delayed pacemaker lead perforations: Why unusual presentations should prompt an early multidisciplinary team approach.

Authors:  John An Kuang Chao; Michael S Firstenberg
Journal:  Int J Crit Illn Inj Sci       Date:  2017 Jan-Mar

5.  An unusual presentation of delayed lead perforation: It's never too late.

Authors:  Shashank Jain; Jude Clancy; Mark H Schoenfeld
Journal:  HeartRhythm Case Rep       Date:  2021-11-26

Review 6.  Non-traditional implantable cardioverter-defibrillator configurations and insertion techniques: a review of contemporary options.

Authors:  Johanna B Tonko; Christopher A Rinaldi
Journal:  Europace       Date:  2022-02-02       Impact factor: 5.214

Review 7.  Time to Shock the System: Moving Beyond the Current Paradigm for Primary Prevention Implantable Cardioverter-Defibrillator Use.

Authors:  Faisal M Merchant; Wayne C Levy; Daniel B Kramer
Journal:  J Am Heart Assoc       Date:  2020-02-24       Impact factor: 5.501

8.  Chest pain following permanent pacemaker insertion… a case of pneumopericardium due to atrial lead perforation.

Authors:  Suleman Aktaa; Kavi Fatania; Claire Gains; Hazel White
Journal:  BMJ Case Rep       Date:  2018-10-25

9.  Delayed presentation of right ventricular lead perforation following defibrillator implantation: a case report.

Authors:  Daniel Ahlert; Andrew R J Mitchell
Journal:  Eur Heart J Case Rep       Date:  2019-07-22

10.  Impact of bedrest on cardiovascular events and complications associated with temporary pacemakers in patients waiting for permanent pacemaker implantation.

Authors:  Tomoaki Hama; Norishige Morita; Akiko Ushijima; Akira Ueno; Takayuki Iida; Yuji Ikari; Yoshinori Kobayashi
Journal:  J Arrhythm       Date:  2021-03-30
  10 in total