Comparison of noninvasive assessments of central blood pressure using general transfer function and late systolic shoulder of the radial pressure wave.

Peter Wohlfahrt1, Alena Krajcoviechová, Jitka Seidlerová, Otto Mayer, Jan Filipovsky, Renata Cífková.   

Abstract

BACKGROUND: Central systolic blood pressure (cSBP) can be derived by the general transfer function of the radial pressure wave, as used in the SphygmoCor device, or by regression equation from directly measured late systolic shoulder of the radial pressure wave (pSBP2), as used in the Omron HEM-9000AI device. The aim of this study was to compare the SphygmoCor estimates of cSBP with 2 estimates of cSBP provided by the Omron HEM-9000AI (cSBP, pSBP2) in a large cohort of the white population.
METHODS: In 391 patients aged 52.3±13.5 years (46% men) from the Czech post-MONICA Study, cSBP was measured using the SphygmoCor and Omron HEM-9000AI devices in random order.
RESULTS: Omron cSBP and pSBP2 were perfectly correlated (r = 1.0; P < 0.0001). There was a strong correlation (r = 0.97; P < 0.0001) between Omron and SphygmoCor cSBP estimates, but Omron estimate was 13.1±4.7mm Hg higher than SphygmoCor cSBP. On the other hand, Omron pSBP2 strongly correlated with SphygmoCor cSBP (r = 0.97; P < 0.0001) and was 1.7±4.2mm Hg lower than SphygmoCor cSBP. In multivariable analysis, anthropometric and cardiovascular risk factors explained only 10% of the variance of the cSBP difference between devices while explaining 52% of the systolic blood pressure amplification variance.
CONCLUSIONS: Estimation of cSBP based on the late systolic shoulder of the radial wave provides a comparable accuracy with the validated general transfer function. When comparing Omron HEM-9000AI and SphygmoCor estimates of cSBP, Omron pSBP2 should be used. The difference between both devices in cSBP may be explained by differences in calibration.

Entities:  

Keywords:  applanation tonometry; blood pressure; central systolic blood pressure; generalized transfer function; hypertension; radial pulse wave; radial second systolic pressure.

Mesh:

Year:  2013        PMID: 23999585     DOI: 10.1093/ajh/hpt166

Source DB:  PubMed          Journal:  Am J Hypertens        ISSN: 0895-7061            Impact factor:   2.689


  5 in total

1.  Central blood pressure estimation by using N-point moving average method in the brachial pulse wave.

Authors:  Rie Sugawara; Shigeo Horinaka; Hiroshi Yagi; Kimihiko Ishimura; Takeharu Honda
Journal:  Hypertens Res       Date:  2015-02-19       Impact factor: 3.872

2.  Clinical Considerations for the Association between Vascular Damage and Chronic Kidney Disease.

Authors:  Hirofumi Tomiyama; Akira Yamashina
Journal:  Pulse (Basel)       Date:  2015-02-27

3.  Arterial Stiffness, Central Pulsatile Hemodynamic Load, and Orthostatic Hypotension.

Authors:  Kai Liu; Si Wang; Shixi Wan; Yufei Zhou; Pei Pan; Bo Wen; Xin Zhang; Hang Liao; Di Shi; Rufeng Shi; Xiaoping Chen; Tulasiram Jangala
Journal:  J Clin Hypertens (Greenwich)       Date:  2015-11-06       Impact factor: 3.738

4.  Joint Effect of Non-invasive Central Systolic Blood Pressure and Peripheral Systolic Blood Pressure on Incident Hypertension in a Chinese Community-based Population.

Authors:  Shixuan Wang; Zechen Zhou; Fangfang Fan; Litong Qi; Jia Jia; Pengfei Sun; Yimeng Jiang; Minghao Kou; Dafang Chen; Yan Zhang; Yong Huo
Journal:  Sci Rep       Date:  2018-02-19       Impact factor: 4.379

5.  Associations between remnant lipoprotein cholesterol and central systolic blood pressure in a Chinese community-based population: a cross-sectional study.

Authors:  Jing Zhou; Yan Zhang; Kaiyin Li; Fangfang Fan; Bo Zheng; Jia Jia; Bo Liu; Jiahui Liu; Chuyun Chen; Yong Huo
Journal:  Lipids Health Dis       Date:  2021-06-26       Impact factor: 3.876

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.