Literature DB >> 23999229

DensiProbe Spine: an intraoperative measurement of bone quality in spinal instrumentation. A clinical feasibility study.

Albrecht W Popp1, Ronald Schwyn, Damiano Schiuma, Marius J Keel, Kurt Lippuner, Lorin M Benneker.   

Abstract

BACKGROUND CONTEXT: A new device, DensiProbe, has been developed to provide surgeons with intraoperative information about bone strength by measuring the peak breakaway torque. In cases of low bone quality, the treatment can be adapted to the patient's condition, for example, by improving screw-anchorage with augmentation techniques.
PURPOSE: The objective of this study was to investigate the feasibility of DensiProbe Spine in patients undergoing transpedicular fixation. STUDY
DESIGN: Prospective feasibility study on consecutive patients. PATIENT SAMPLE: Fourteen women and 16 men were included in this study. OUTCOME MEASURES: Local and general bone quality.
METHODS: These consecutive patients scheduled for transpedicular fixation were evaluated for bone mineral density (BMD), which was measured globally by dual-energy X-ray absorptiometry and locally via biopsies using quantitative microcomputed tomography. The breakaway torque force within the vertebral body was assessed intraoperatively via the transpedicular approach with the DensiProbe Spine. The results were correlated with the areal BMD at the lumbar spine and the local volumetric BMD (vBMD) and a subjective impression of bone strength. The feasibility of the method was evaluated, and the clinical and radiological performance was evaluated over a 1-year follow-up. This study was funded by an AO Spine research grant; DensiProbe was developed at the AO Research Institute Davos, Switzerland; the AO Foundation is owner of the intellectual property rights.
RESULTS: In 30 patients, 69 vertebral levels were examined. The breakaway torque consistently correlated with an experienced surgeon's quantified impression of resistance as well as with vBMD of the same vertebra. Beyond a marginal prolongation of surgery time, no adverse events related to the usage of the device were observed.
CONCLUSIONS: The intraoperative transpedicular measurement of the peak breakaway torque was technically feasible, safe, and reliably predictive of local vBMD during dorsal spinal instrumentations in a clinical setting. Larger studies are needed to define specific thresholds that indicate a need for the augmentation or instrumentation of additional levels.
Copyright © 2013 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Bone mineral density; Local bone strength; Pedicle screw failure; Spinal fixation

Mesh:

Year:  2013        PMID: 23999229     DOI: 10.1016/j.spinee.2013.06.067

Source DB:  PubMed          Journal:  Spine J        ISSN: 1529-9430            Impact factor:   4.166


  4 in total

1.  Technologies for assessment of bone reflecting bone strength and bone mineral density in elderly women: an update.

Authors:  Alvilde Dhainaut; Mari Hoff; Unni Syversen; Glenn Haugeberg
Journal:  Womens Health (Lond)       Date:  2016-02-22

2.  Local bone quality measure and construct failure prediction: a biomechanical study on distal femur fractures.

Authors:  Thomas Vordemvenne; Dirk Wähnert; Dominic Gehweiler; Ursula Styger; Boyko Gueorguiev; Christian Colcuc
Journal:  Arch Orthop Trauma Surg       Date:  2021-02-15       Impact factor: 2.928

3.  Assessment of the Breakaway Torque at the Posterior Pelvic Ring in Human Cadavers.

Authors:  Johannes Dominik Bastian; Mathias Bergmann; Ronald Schwyn; Marius Johann Baptist Keel; Lorin Michael Benneker
Journal:  J Invest Surg       Date:  2015-08-13       Impact factor: 2.533

4.  Preoperative dual-energy X-ray absorptiometry and FRAX in patients with lumbar spinal stenosis.

Authors:  Camilla Bergh; Ann-Charlott Söderpalm; Helena Brisby
Journal:  J Orthop Surg Res       Date:  2018-10-16       Impact factor: 2.359

  4 in total