Conductor externalization of the RIATA implantable cardioverter defibrillator lead: a challenging lead extraction requiring laser-powered sheath.

Federico Migliore1, Emanuele Bertaglia, Luca Bontempi, Roberto Verlato, Loira Leoni, Sabino Iliceto, Antonio Curnis.   

Abstract

We report the case of a 27-year-old man who received a single chamber implantable cardioverter defibrillator with a St Jude Medical Riata double coil passive lead for secondary prevention of sudden cardiac death due to idiopathic ventricular fibrillation. Subsequently, he was readmitted for lead extraction because of device infection. Fluoroscopy showed externalization of the conductors proximally to the distal coil in the absence of any signs of electrical lead failure. Successful lead extraction was achieved using a laser sheath without any complications after unsuccessful mechanical sheath because of the tenacious fibrotic adherences proximal to the distal coil and extrusion of the conductors that hindered the advancement of the sheath over the lead. Our report highlights that Riata ICD leads are prone to externalization of the conductor wires, which may render an indicated lead extraction procedure of these leads more challenging. A laser powered sheath may facilitate the procedure, especially in the presence of extensive adhesions and fibrosis.

Entities:  

Mesh:

Year:  2013        PMID: 23996009     DOI: 10.2459/JCM.0b013e3283628c8d

Source DB:  PubMed          Journal:  J Cardiovasc Med (Hagerstown)        ISSN: 1558-2027            Impact factor:   2.160


  1 in total

1.  Combined LASER and femoral approach to remove a previous failure of Riata lead extraction.

Authors:  Antonio Dello Russo; Gaetano Michele Fassini; Sergio Conti; Giulia Vettor; Michela Casella; Fabrizio Tundo; Maurizio Roberto; Francesco Grillo; Francesco Alamanni; Claudio Tondo
Journal:  Clin Case Rep       Date:  2017-07-20
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.