Literature DB >> 23995144

Direct beam radiation exposure to surgeons during pinning of supracondylar humerus fractures: does C-arm position and the attending surgeon matter?

Emily A Eismann1, Eric J Wall, Elizabeth C Thomas, Megan A Little.   

Abstract

BACKGROUND: Direct beam radiation exposure to the surgeon, especially to their hands, is extremely common during supracondylar humerus fracture pinnings and results in exposure to significantly greater doses of ionizing radiation when compared with scatter radiation. The purpose of this study was to determine how often surgeons are exposed to direct beam radiation during this surgery and whether the C-arm position and the surgeon's experience influence radiation exposure.
METHODS: In this double blind study, we collected 3842 fluoroscopic still images from 78 closed reduction and percutaneous pinning surgeries for supracondylar humerus fractures performed or supervised by 6 attending surgeons. The percentage of images containing a surgeon's body was calculated as an indicator of direct beam radiation exposure. Total fluoroscopy time, C-arm position (standard or inverted), and whether the primary surgeon was an attending, resident, or both were recorded. Nonparametric statistical analyses were performed.
RESULTS: Fluoroscopy lasted for a median of 34 seconds, and the surgeon was exposed to direct beam radiation in a median of 13% of fluoroscopy films, with exposure ranging from 0% to 97% per surgery. Fluoroscopy was significantly longer when the C-arm position was inverted when compared with the standard position (43 vs. 26 s, P=0.034). Surgeons' exposure to direct beam radiation was also slightly greater when the C-arm position was inverted (16% vs. 10%, P=0.087). The duration of fluoroscopy exposure and the percentage of films with the body exposed to radiation did not differ based on whether the surgery was performed by an attending, a resident, or both (P=0.53 and 0.28, respectively). However, the percentage of films with bodily radiation exposure did significantly differ between the attending physicians (P=0.029).
CONCLUSIONS: Direct beam radiation exposure varied widely between surgeries and surgeons, ranging from none to nearly constant exposure. Surgical time also significantly increased with the C-arm in the inverted position compared with the standard position. Given the significant variation in exposure between attending physicians, it is likely that exposure to direct beam radiation can be avoided with improved awareness about the risk of direct beam radiation exposure and cautious surgical technique. LEVEL OF EVIDENCE: Not applicable.

Entities:  

Mesh:

Year:  2014        PMID: 23995144     DOI: 10.1097/BPO.0000000000000086

Source DB:  PubMed          Journal:  J Pediatr Orthop        ISSN: 0271-6798            Impact factor:   2.324


  3 in total

1.  Quantification of radiation exposure in the operating theatre during management of common fractures of the upper extremity in children.

Authors:  J F Maempel; O D Stone; A W Murray
Journal:  Ann R Coll Surg Engl       Date:  2016-09       Impact factor: 1.891

2.  Real-time, ray casting-based scatter dose estimation for c-arm x-ray system.

Authors:  Zaid Alnewaini; Eric Langer; Philipp Schaber; Matthias David; Dominik Kretz; Volker Steil; Jürgen Hesser
Journal:  J Appl Clin Med Phys       Date:  2017-01-24       Impact factor: 2.102

3.  A NEW FLUOROSCOPY TECHNIQUE FOR SUPRACONDYLAR HUMERUS FRACTURES.

Authors:  Mirza Zafer Dagtas; Omer Kays Unal
Journal:  Acta Ortop Bras       Date:  2022-01-28       Impact factor: 0.513

  3 in total