| Literature DB >> 23993215 |
Wonmok Lee1, Hae-Sun Chung, Yangsoon Lee, Dongeun Yong, Seok Hoon Jeong, Kyungwon Lee, Yunsop Chong.
Abstract
Matrix-assisted laser desorption ionization-time-of-flight mass spectrometry assay was able to detect carbapenemase producers, including SIM-1 or OXA-51, within 4 hours using 20 μL of 0.5 g/L ertapenem solution as a substrate. This assay is more rapid and accurate than the modified Hodge test and 3-dimensional extract bioassay. Hence, it can be used an alternative test to identify carbapenemase-mediated carbapenem resistance in Gram-negative bacteria.Entities:
Keywords: Acinetobacter spp; Carbapenemase; MALDI-TOF MS; OXA; SIM-1
Mesh:
Substances:
Year: 2013 PMID: 23993215 DOI: 10.1016/j.diagmicrobio.2013.07.005
Source DB: PubMed Journal: Diagn Microbiol Infect Dis ISSN: 0732-8893 Impact factor: 2.803