Hydrophilic guidewires: evaluation and comparison of their properties and safety.

Fabio Cesar Miranda Torricelli1, Shubha De, Carl Sarkissian, Manoj Monga.   

Abstract

OBJECTIVE: To compare physical and mechanical properties of 10 commercially available hydrophilic guidewires.
METHODS: In vitro testing was performed to evaluate 10 different straight hydrophilic guidewires (5 regular and 5 stiff wires): Glidewire, NiCore, EZ Glider, Hiwire, and Zipwire. The forces required for tip perforation, tip bending, shaft bending, and friction during movement were measured for all 10 wires. The tip contour was measured using high power light microscopy.
RESULTS: The Glidewire required the greatest force to perforate our model (P = .01). The EZ Glider, Zipwire, and Glidewire had the lowest tip bending forces (P <.001). The Glidewire had the stiffest shaft (P <.001). The EZ Glider and Glidewire required the greatest forces in the friction test (P <.001). Regarding the stiff guidewires, the GlidewireS required the greatest force in the perforation test (P ≤.05). The GlidewireS and EZ GliderS required the lowest tip bending force (P ≤.004). The ZipwireS and NiCoreS had the stiffest shafts (P ≤.01). The GlidewireS required the greatest force in the friction test (P <.001). Measurement of the tip contour showed the Zipwire, HiwireS, and EZ GliderS had the roundest tips.
CONCLUSION: Each wire has unique properties with advantages and disadvantages. The Glidewires (both stiff and regular) have the lowest potential for perforation, although they are less slippery. The Glidewire and EZ Glider required the least tip force to bend around a point of obstruction.
Copyright © 2013 Elsevier Inc. All rights reserved.

Mesh:

Year:  2013        PMID: 23992968     DOI: 10.1016/j.urology.2013.07.024

Source DB:  PubMed          Journal:  Urology        ISSN: 0090-4295            Impact factor:   2.649


  3 in total

1.  Comparative study of wire-integrated urethral catheter versus wire-assisted urethral catheter in difficult male urethral catheterization: a single-center experience.

Authors:  Peng He; Siji Song; Zhipeng Chen; Yang Liu; Zhansong Zhou; Yongquan Wang
Journal:  Transl Androl Urol       Date:  2019-12

2.  Ureteral stents should be soaked for several minutes before placement.

Authors:  Norbert Laube; Chintan Desai; Falk Bernsmann; Christian Fisang
Journal:  Springerplus       Date:  2015-06-09

3.  Are All Wires Created the Same? A Quality Assurance Study of the Stiffness of Wires Typically Employed During Endovascular Surgery Using Tension Dynamometry.

Authors:  Arindam Chaudhuri; Frederic Heim; Nabil Chakfe
Journal:  EJVES Vasc Forum       Date:  2021-06-22
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.