Literature DB >> 2397260

Deterioration mechanisms of joint prosthesis materials. Several solutions by ion implantation surface treatments.

J Rieu1, A Pichat, L M Rabbe, C Chabrol, M Robelet.   

Abstract

Materials for orthopaedic implants can fail for several combined reasons: corrosion, fatigue and wear for metals, wear and creep for polymers, fracture for ceramics. Some typical cases are analysed and it is demonstrated that ion implantation improves metals and polymers used for joint prosthesis. Implantations of nitrogen, oxygen and argon ions modify the structure of a 2-500 nm thick layer in the materials. The results of friction tests on the couple metal-polymer are correlated with the surface properties.

Entities:  

Mesh:

Substances:

Year:  1990        PMID: 2397260

Source DB:  PubMed          Journal:  Biomaterials        ISSN: 0142-9612            Impact factor:   12.479


  1 in total

1.  In vivo evaluation of bone tissue behavior on ion implanted surfaces.

Authors:  M Bosetti; A Massè; E Tobin; M Cannas
Journal:  J Mater Sci Mater Med       Date:  2001-05       Impact factor: 3.896

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.