Literature DB >> 23972313

Intraocular pressure fluctuation during microincision vitrectomy with constellation vision system.

Yoshimi Sugiura1, Fumiki Okamoto, Yoshifumi Okamoto, Takahiro Hiraoka, Tetsuro Oshika.   

Abstract

PURPOSE: To investigate intraocular pressure (IOP) fluctuation during various vitrectomy maneuvers using the vitrectomy system (Alcon Constellation Vision System).
DESIGN: An experimental study as laboratory investigation.
METHODS: In porcine eyes, 23- and 25-gauge vitrectomy was performed, and IOP fluctuations were evaluated in vitreous cutting mode, in aspiration mode, and during scleral compression. The measurements were performed with the IOP control setting turned on or off.
RESULTS: Using the 23-gauge system with the IOP control setting turned on, IOP decreased from 30 to 23.7 mm Hg after starting vitreous cutting, and then returned to 30 mm Hg in 2.6 seconds. When the IOP control setting was turned off, IOP decreased to 19.1 mm Hg in 0.9 seconds, and remained at that pressure. Under aspiration at 650 mm Hg without cutting, IOP showed a sharp depression from 30 to 12.2 mm Hg, and then returned to 30.6 mm Hg in 2.6 seconds with the IOP control setting turned on. When the IOP control setting was turned off, IOP decreased to 2.2 mm Hg in 9.7 seconds, and did not recover. When the sclera was compressed without aspiration, IOP rapidly increased to 70-100 mm Hg, and then slowly decreased to 30 mm Hg in 3.5-4.0 seconds, with or without the IOP control system. Similar data were obtained with 25-gauge vitrectomy.
CONCLUSIONS: The IOP control system can attenuate IOP fluctuations during vitrectomy maneuvers. There was no significant difference in IOP fluctuations between 23- and 25-gauge systems.
Copyright © 2013 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23972313     DOI: 10.1016/j.ajo.2013.06.016

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  6 in total

1.  Dynamic Pressure Measurements During Vitrectomy in a Model of the Eye.

Authors:  Irene Nepita; Alessandro Stocchino; Andrea Dodero; Maila Castellano; Mariantonia Ferrara; Mario R Romano; Rodolfo Repetto
Journal:  Transl Vis Sci Technol       Date:  2022-05-02       Impact factor: 3.048

2.  Effect of intraocular lens diameter implanted in enucleated porcine eye on intraocular pressure induced by scleral depression.

Authors:  Gaku Terauchi; Celso Soiti Matsumoto; Kei Shinoda; Harue Matsumoto; Atsushi Mizota
Journal:  Biomed Res Int       Date:  2014-03-27       Impact factor: 3.411

3.  Intraoperative electrophysiological evaluations of macular function during peripheral scleral indentation.

Authors:  Goichi Akiyama; Celso Soiti Matsumoto; Kei Shinoda; Gaku Terauchi; Harue Matsumoto; Emiko Watanabe; Takeshi Iwata; Atsushi Mizota; Yozo Miyake
Journal:  Sci Rep       Date:  2016-10-20       Impact factor: 4.379

4.  Electroretinographic evaluations of retinal function before, just after, and after intravitreal injections.

Authors:  Kazuma Yagura; Kei Shinoda; Soiti Matsumoto; Gaku Terauchi; Makoto Kawashima; Emiko Watanabe; Harue Matsumoto; Takeshi Iwata; Atsushi Mizota; Yozo Miyake
Journal:  Sci Rep       Date:  2016-08-05       Impact factor: 4.379

5.  Jet stream related iatrogenic retinal breaks during vitreo-retinal surgery.

Authors:  Akash Belenje; Brijesh Takkar; Komal Agarwal; Mudit Tyagi; Vinod Aggarwal; Tapas R Padhi; Raja Narayanan
Journal:  Indian J Ophthalmol       Date:  2022-03       Impact factor: 2.969

6.  Commentary: Iatrogenic retinal breaks secondary to jet stream injury.

Authors:  Pranjal Mishra; Harshita Nahata; Naresh K Yadav; Ramesh Venkatesh
Journal:  Indian J Ophthalmol       Date:  2022-03       Impact factor: 1.848