Literature DB >> 23965825

The feasibility of non-contact ultrasound for medical imaging.

G T Clement1, H Nomura, H Adachi, T Kamakura.   

Abstract

High intensity focused ultrasound in air may provide a means for medical and biological imaging without direct coupling of an ultrasound probe. In this study, an approach based on highly focused ultrasound in air is described and the feasibility of the technique is assessed. The overall method is based on the observations that (1) ultrasound in air has superior focusing ability and stronger nonlinear harmonic generation as compared to tissue propagation and (2) a tightly focused field directed into tissue causes point-like spreading that may be regarded as a source for generalized diffraction tomography. Simulations of a spherically-curved transducer are performed, where the transducer's radiation pattern is directed from air into tissue. It is predicted that a focal pressure of 162 dB (2.5 kPa) is sufficient to direct ultrasound through the body, and provide a small but measurable signal (∼1 mPa) upon exit. Based on the simulations, a 20 cm diameter array consisting of 298 transducers is constructed. For this feasibility study, a 40 kHz resonance frequency is selected based on the commercial availability of such transducers. The array is used to focus through water and acrylic phantoms, and the time history of the exiting signal is evaluated. Sufficient data are acquired to demonstrate a low-resolution tomographic reconstruction. Finally, to demonstrate the feasibility to record a signal in vivo, a 75 mm × 55 mm section of a human hand is imaged in a C-mode configuration.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23965825     DOI: 10.1088/0031-9155/58/18/6263

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  1 in total

1.  Ultrasound field measurement using a binary lens.

Authors:  Gregory Clement; Hideyuki Nomura; Tomoo Kamakura
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2015-02       Impact factor: 2.725

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.