Results at a minimum of 10 years of follow-up for AMS and PerFix HA-coated cementless total hip arthroplasty: impact of cross-linked polyethylene on implant longevity.

Yasuharu Nakashima1, Taishi Sato, Takuaki Yamamoto, Goro Motomura, Masanobu Ohishi, Satoshi Hamai, Mio Akiyama, Masanobu Hirata, Daisuke Hara, Yukihide Iwamoto.   

Abstract

PURPOSE: The purpose of this study was to report results at a minimum of 10 years for hydroxyapatite (HA) coating on the titanium arc-sprayed cementless total hip arthroplasty (THA) and to evaluate the impact of cross-linked polyethylene (XLPE) on implant longevity.
METHODS: A total of 131 consecutive primary THAs in 123 patients using an AMS acetabular cup and a PerFix HA stem were retrospectively reviewed. Conventional PE was used for 62 hips (CPE group) and cross-linked PE for 69 hips (XLPE group), with mean follow-up periods of 13 and 11.5 years, respectively. These patients were reviewed using the Japanese Orthopaedic Association (JOA) Hip Score and evaluated in terms of PE wear, osteolysis, and implant survival.
RESULTS: The JOA score improved from 42.6 to 83.9 points at the final follow-up. The mean wear rate of 0.12 mm/year in the CPE group was significantly greater than that of XLPE at 0.007 mm/year. In the CPE group, ten (16.1 %) and two (3.2 %) hips, respectively, underwent PE exchange due to severe wear and acetabular revision due to loosening associated with osteolysis. Three patients had revision in the XLPE group: one cup for aseptic loosening, one PE for recurrent dislocation, and one stem for neck fracture. No evident osteolysis was seen in the XLPE group. Kaplan-Meier survivorship with any revision as the end point shows that the 12-year survival rate was 97.7 % for cups, 93.8 % for PE liners, and 99.2 % for stems. Multivariate analysis revealed that the use of XLPE significantly reduced the risk of revision, with the odds ratio (OR) of 4.3.
CONCLUSIONS: AMS and PerFix HA components in this series show excellent implant fixation; however, high rates of PE wear and subsequent osteolysis were limiting factors in long-term success. Low wear rates with XLPE suggest improved implant longevity.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23963589     DOI: 10.1007/s00776-013-0456-4

Source DB:  PubMed          Journal:  J Orthop Sci        ISSN: 0949-2658            Impact factor:   1.601


  18 in total

1.  Is there a difference in total knee arthroplasty risk of revision in highly crosslinked versus conventional polyethylene?

Authors:  Elizabeth W Paxton; Maria C S Inacio; Steven Kurtz; Rebecca Love; Guy Cafri; Robert S Namba
Journal:  Clin Orthop Relat Res       Date:  2015-03       Impact factor: 4.176

2.  Metal-on-conventional polyethylene total hip arthroplasty bearing surfaces have a higher risk of revision than metal-on-highly crosslinked polyethylene: results from a US registry.

Authors:  Elizabeth W Paxton; Maria C S Inacio; Robert S Namba; Rebecca Love; Steven M Kurtz
Journal:  Clin Orthop Relat Res       Date:  2015-03       Impact factor: 4.176

3.  Dislocation and its recurrence after revision total hip arthroplasty.

Authors:  Kensei Yoshimoto; Yasuharu Nakashima; Takuaki Yamamoto; Jun-Ichi Fukushi; Goro Motomura; Masanobu Ohishi; Satoshi Hamai; Yukihide Iwamoto
Journal:  Int Orthop       Date:  2015-10-20       Impact factor: 3.075

4.  Optimal anterior femoral offset for functional range of motion in total hip arthroplasty--a computer simulation study.

Authors:  Masanobu Hirata; Yasuharu Nakashima; Daisuke Hara; Masayuki Kanazawa; Yusuke Kohno; Kensei Yoshimoto; Yukihide Iwamoto
Journal:  Int Orthop       Date:  2014-09-25       Impact factor: 3.075

5.  Long-term clinical outcome of acetabular cup revision surgery: comparison of cemented cups, cementless cups, and cemented cups with reinforcement devices.

Authors:  Yasuo Kokubo; Hisashi Oki; Daisuke Sugita; Kohei Negoro; Kenichi Takeno; Tsuyoshi Miyazaki; Hideaki Nakajima
Journal:  Eur J Orthop Surg Traumatol       Date:  2016-03-24

6.  Re-dislocation after revision total hip arthroplasty for recurrent dislocation: a multicentre study.

Authors:  Kensei Yoshimoto; Yasuharu Nakashima; Shigeo Aota; Ayumi Kaneuji; Kiyokazu Fukui; Kazuo Hirakawa; Nariaki Nakura; Koichi Kinoshita; Masatoshi Naito; Yukihide Iwamoto
Journal:  Int Orthop       Date:  2016-02-19       Impact factor: 3.075

7.  Gait instability may indicate liner failure in patients with total hip arthroplasty. A report of three cases.

Authors:  F Díaz-Dilernia; M Lattore; G Zanotti; F Comba; F Piccaluga; M Buttaro
Journal:  Ann R Coll Surg Engl       Date:  2021-08-20       Impact factor: 1.951

8.  Cementless total hip arthroplasty for patients previously treated with femoral osteotomy for hip dysplasia: the incidence of periprosthetic fracture.

Authors:  Masanobu Ohishi; Yasuharu Nakashima; Takuaki Yamamoto; Goro Motomura; Jun-Ichi Fukushi; Satoshi Hamai; Yusuke Kohno; Yukihide Iwamoto
Journal:  Int Orthop       Date:  2015-09-19       Impact factor: 3.075

9.  Cementless hydroxyapatite coated hip prostheses.

Authors:  Antonio Herrera; Jesús Mateo; Jorge Gil-Albarova; Antonio Lobo-Escolar; Elena Ibarz; Sergio Gabarre; Yolanda Más; Luis Gracia
Journal:  Biomed Res Int       Date:  2015-02-23       Impact factor: 3.411

10.  Total Hip Arthroplasty for the Paralytic and Non-paralytic Side in Patient with Residual Poliomyelitis.

Authors:  Motoki Sonohata; Masaru Kitajima; Shunsuke Kawano; Masaaki Mawatari
Journal:  Open Orthop J       Date:  2016-05-13
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.