Literature DB >> 23958073

A comparison of the EZ-Blocker with a Cohen Flex-Tip blocker for one-lung ventilation.

Alparslan Kus1, Tulay Hosten2, Yavuz Gurkan2, Aslı Gul Akgul3, Mine Solak2, Kamil Toker2.   

Abstract

OBJECTIVES: The EZ-Blocker (IQ Medical Ventures BV, Rotterdam, Netherlands) is a newly designed device for one-lung ventilation. The aim of this study was to compare the effectiveness of the Cohen Flex-Tip bronchial blocker (Cook, Bloomington, IN) and the EZ-Blocker for one-lung ventilation during thoracic surgery.
DESIGN: Randomized and prospective.
SETTING: A university hospital. PARTICIPANTS: This study included 40 patients undergoing thoracic surgical procedures.
INTERVENTIONS: Patients were assigned to 2 study groups: Patients who received the Cohen Flex-Tip blocker were assigned to the Cohen group, and patients who received the EZ-Blocker were assigned to the EZ group. In both groups, fiberoptic guidance was used during placement of the bronchial blockers. Comparisons between the groups included the time to correct placement, the incidence of malpositioning, and the satisfaction level of the surgeon (good, fair, poor).
MEASUREMENTS AND MAIN RESULTS: One-lung ventilation was achieved successfully for all patients. The time to correct placement (mean±SD) was significantly shorter in the EZ group (146±56 seconds) compared with the Cohen group (241±51 seconds; p=0.01). The incidence of malpositioning was significantly lower in the EZ group compared with the Cohen group (p=0.018). Surgeon satisfaction was similar in both groups.
CONCLUSIONS: In this study, both bronchial blockers provided similar surgical exposure during thoracic procedures. The EZ-Blocker had a shorter time to correct positioning and less frequent intraoperative malpositioning.
Copyright © 2014 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Cohen Flex-Tip bronchial blocker; EZ-Blocker; bronchial blocker; one-lung ventilation

Mesh:

Year:  2013        PMID: 23958073     DOI: 10.1053/j.jvca.2013.02.006

Source DB:  PubMed          Journal:  J Cardiothorac Vasc Anesth        ISSN: 1053-0770            Impact factor:   2.628


  4 in total

Review 1.  [Airway management for lung separation in thoracic surgery : An update].

Authors:  K M Meggiolaro; H Wulf; C Feldmann; T Wiesmann; A-K Schubert; J Risse
Journal:  Anaesthesist       Date:  2018-08       Impact factor: 1.041

2.  Use of Fogarty catheter as bronchial blocker for lung isolation in children undergoing thoracic surgery: A single centre experience of 15 cases.

Authors:  Bikram K Behera; Satyajeet Misra; Manoj K Mohanty; Bikasha B Tripathy
Journal:  Ann Card Anaesth       Date:  2022 Apr-Jun

3.  The EZ-blocker for one-lung ventilation in patients undergoing thoracic surgery: clinical applications and experience in 100 cases in a routine clinical setting.

Authors:  Andreas Moritz; Andrea Irouschek; Torsten Birkholz; Johannes Prottengeier; Horia Sirbu; Joachim Schmidt
Journal:  J Cardiothorac Surg       Date:  2018-06-25       Impact factor: 1.637

4.  Comparison of left double lumen tube and y-shaped and double-ended bronchial blocker for one lung ventilation in thoracic surgery-a randomised controlled clinical trial.

Authors:  Joachim Risse; Karsten Szeder; Ann-Kristin Schubert; Thomas Wiesmann; Hanns-Christian Dinges; Carsten Feldmann; Hinnerk Wulf; Karl Matteo Meggiolaro
Journal:  BMC Anesthesiol       Date:  2022-04-02       Impact factor: 2.217

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.