Dosimetric verification of complex radiotherapy with a 3D optically based dosimetry system: dose painting and target tracking.

Peter S Skyt1, Jørgen B B Petersen, Esben S Yates, Per R Poulsen, Thomas L Ravkilde, Peter Balling, Ludvig P Muren.   

Abstract

BACKGROUND: The increasing complexity of radiotherapy (RT) has motivated research into three-dimensional (3D) dosimetry. In this study we investigate the use of 3D dosimetry with polymerizing gels and optical computed tomography (optical CT) as a verification tool for complex RT: dose painting and target tracking.
MATERIALS AND METHODS: For the dose painting studies, two dosimeters were irradiated with a seven-field intensity modulated radiotherapy (IMRT) plan with and without dose prescription based on a hypoxia image dataset of a head and neck patient. In the tracking experiments, two dosimeters were irradiated with a volumetric modulated arc therapy (VMAT) plan with and without clinically measured prostate motion and a third with both motion and target tracking. To assess the performance, 3D gamma analyses were performed between measured and calculated stationary dose distributions.
RESULTS: Gamma pass-rates of 95.3% and 97.3% were achieved for the standard and dose-painted IMRT plans. Gamma pass-rates of 91.4% and 54.4% were obtained for the stationary and moving dosimeter, respectively, while tracking increased the pass-rate for the moving dosimeter to 90.4%.
CONCLUSIONS: This study has shown that the 3D dosimetry system can reproduce and thus verify complex dose distributions, also when influenced by motion.

Entities:  

Mesh:

Year:  2013        PMID: 23957684     DOI: 10.3109/0284186X.2013.813965

Source DB:  PubMed          Journal:  Acta Oncol        ISSN: 0284-186X            Impact factor:   4.089


  3 in total

1.  Dosimetric and Radiobiological Evaluation of Multiparametric MRI-Guided Dose Painting in Radiotherapy of Prostate Cancer.

Authors:  Iraj Abedi; Mohammad B Tavakkoli; Keyvan Jabbari; Alireza Amouheidari; Ghasem Yadegarfard
Journal:  J Med Signals Sens       Date:  2017 Apr-Jun

Review 2.  Three-dimensional radiation dosimetry using polymer gel and solid radiochromic polymer: From basics to clinical applications.

Authors:  Yoichi Watanabe; Leighton Warmington; N Gopishankar
Journal:  World J Radiol       Date:  2017-03-28

3.  Empirical quenching correction in radiochromic silicone-based three-dimensional dosimetry of spot-scanning proton therapy.

Authors:  Lia Barbosa Valdetaro; Ellen Marie Høye; Peter Sandegaard Skyt; Jørgen Breede Baltzer Petersen; Peter Balling; Ludvig Paul Muren
Journal:  Phys Imaging Radiat Oncol       Date:  2021-04-12
  3 in total