Do ultrasonic dental scalers interfere with implantable cardioverter defibrillators? An in vivo investigation.

Carlo Maiorana1, Giovanni B Grossi, Rocco A Garramone, Ruggero Manfredini, Franco Santoro.   

Abstract

OBJECTIVES: To test the in vivo effects of an ultrasonic dental scaler on various implanted cardioverter defibrillator (ICD) models.
METHODS: 12 consecutive patients with ICDs had continuous both electrocardiogram monitoring and device interrogation to detect interferences during the use of an ultrasonic dental scaler.
RESULTS: No interferences were detected by any ICD. Evaluation of the electrocardiograms for each patient failed to show any abnormalities in pacing during testing.
CONCLUSION: The results of this study suggest that the routinary clinic use of piezoelectric dental scalers do not interfere with the functioning of any of the tested ICDs. CLINICAL SIGNIFICANCE: Ultrasonic dental scalers have been suspected of electromagnetic interference (EMI) with the normal functioning of ICDs and the use of this type of equipment for patients with these devices has been controversial. This is the first in vivo study to investigate EMI of ICD activity during the operation with ultrasonic dental scaler.
Copyright © 2013 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Electromagnetic interference; Implantable cardioverter defibrillator; Ultrasonic dental scaler

Mesh:

Year:  2013        PMID: 23948395     DOI: 10.1016/j.jdent.2013.08.006

Source DB:  PubMed          Journal:  J Dent        ISSN: 0300-5712            Impact factor:   4.379


  2 in total

Review 1.  Management of Patients With Cardiovascular Implantable Electronic Devices in Dental, Oral, and Maxillofacial Surgery.

Authors:  James Tom
Journal:  Anesth Prog       Date:  2016

2.  Electromagnetic interference in cardiac electronic implants caused by novel electrical appliances emitting electromagnetic fields in the intermediate frequency range: a systematic review.

Authors:  Sarah Driessen; Andreas Napp; Kristina Schmiedchen; Thomas Kraus; Dominik Stunder
Journal:  Europace       Date:  2019-02-01       Impact factor: 5.214

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.